Performance Testing and Calibration

4

Calibration

3.From the CAL directory on the PC, type CAL. Then press any key to start the program and access the SETUP menu.

4.On Hydra, press POWER ON. After the initialization process has concluded, use the following procedure to set up communications:

a.Press SHIFT and then LIST(COMM).

b.With ’BAUd’ displayed, use the UP or DOWN arrow key to selectthe desired baud rate. Then press ENTER.

c.With ’PAR’ (parity) displayed, use the UP or DOWN arrow key toselect the parity. Then press ENTER.

d.With ’CtS’ (2635A only; Clear to Send) displayed, use the UP orDOWN arrow key to select the Clear to Send flow control ’OFF’.Then press ENTER.

e.With ’ECHO’ displayed, use the UP or DOWN arrow key to select’OFF’. Then press ENTER. Communications setup for Hydra is nowcomplete.

5.On the PC, use the SETUP menu to match the communication parameters defined above for Hydra.

6.On Hydra, break the calibration seal on the front panel display. Then press and hold the CAL Enable button (approximately 4 seconds) until ’CAL’ is displayed. Press this button with a blunt-tipped object. Avoid using a sharper-tipped object (such as a pencil).

7.On the PC, use the right and left arrow keys to select CAL. Then press the ENTER key. A message asking if you want to calibrate is displayed. Press Y and ENTER. The next displayed message specifies the voltage to be applied to channel 1.

4-23. Calibration Procedure Using Starter

Use the following procedure to calibrate Hydra with the Hydra Starter Package:

1.Connect the channel 1 test leads to the 5700A output.

2.On the 5700A, select the output voltage specified on the PC (step 7 above.)

3.On the PC, press ENTER. If the input voltage is within a predetermined acceptable boundary, Hydra performs a calibration for this step. The program then prompts you for the next input value.

Note

"Bad Calibration Input Value" is returned if the input is not acceptable (the calibration step could not be executed.) Verify that the input to Hydra channel 1 is the correct value and polarity. Also verify that the 5700A is in OPERATE mode. If the input is correct and "Bad Calibration Input Value" is still returned, repair of Hydra may be required.

4.Following the prompts, complete all steps for this calibration group.

5.You will then be asked if you want to perform the next calibration group. Press Y -

ENTER.

6.Following the prompts, complete all steps in the remaining calibration groups.

4-21

Page 161
Image 161
Fluke 2625A, 2635A service manual Calibration Procedure Using Starter

2635A, 2625A specifications

The Fluke 2625A is a sophisticated data logger designed for a wide range of industrial and laboratory applications. This versatile instrument excels in capturing and monitoring temperature and voltage measurements with precision and reliability. One of its standout features is its ability to record data from various sensors, making it ideal for environments where comprehensive data acquisition is critical.

Equipped with a large LCD display, the Fluke 2625A provides clear and user-friendly access to real-time measurement data. Its graphical interface allows users to visualize trends and monitor readings easily, which significantly enhances usability. The device supports a variety of input types and can connect to thermocouples, RTDs, and other sensors, offering great flexibility for various measurement tasks.

The Fluke 2625A employs advanced measurement technologies that ensure accuracy in data logging. With a high sampling rate, it captures fast-changing signals while maintaining data integrity. The device features a built-in memory that allows for extensive logging, accommodating long-term monitoring tasks without requiring constant oversight. Users can set it to log data at specified intervals, offering customizable solutions for different monitoring needs.

Another significant characteristic of the Fluke 2625A is its robust communication capabilities. The device supports USB and RS-232 interfaces for easy connectivity with computers and other devices, making data transfer and analysis seamless. This feature is especially beneficial for users who need to compile or analyze large sets of data quickly.

Furthermore, the Fluke 2625A is designed with durability in mind. Its rugged construction makes it suitable for challenging environments, including those with high levels of vibration, dust, or moisture. This reliability ensures that the device can withstand the rigors of fieldwork while still delivering accurate measurements.

In summary, the Fluke 2625A is a powerful data logger that combines versatility, accuracy, and reliability. Its user-friendly interface, advanced measurement technologies, and robust design make it an essential tool for professionals in various industries, from manufacturing and HVAC to research and development. Whether monitoring temperature changes in a laboratory or assessing voltage levels in an industrial setting, the Fluke 2625A stands out as a trusted solution for effective data logging and analysis.