2-47. Totalizer Input

The Totalizer Input circuit consists of Input Protection, a Digital Input Buffer circuit, and a Totalizer Debouncer circuit. The Digital Input Buffer for the totalizer is protected from electrostatic discharge (ESD) damage by A1R49 and A1C43. Refer to the detailed description of the Digital Input Buffer circuit for more information.

The Totalizer Debounce circuit allows the Microprocessor to select totalizing of either the input signal or the debounced input signal. Latch output A1U16-16 is set low by A1U4 to totalize the unmodified input signal or high to totalize the debounced input signal. This totalizer clock control is provided by A1U28; output A1U28-3 drives the totalizer counter clock input (A1U2-4).

The actual debouncing of the input signal is accomplished by A1U14, A1U20, and A1U29. An EXOR gate compares the input signal (A1U14-13) and the output of an eight-bit shift register (A1U29-9). If these signals differ, EXOR gate output A1U14-11 goes high, enabling counter A1U20 and shift register A1U29. The counter divides the system clock of 1.2288 MHz (A1U20-10) by 256 to yield a 4.8-kHz clock (A1U20-13). This signal clocks the eight-bit shift register. After approximately 1.5625 milliseconds, the input signal will have been shifted from the serial input (A1U29-10) through to the eighth output bit (A1U29-9). This forces the counter and shift register to stop. If the input signal changes state before 1.5625 milliseconds have elapsed, the counter is cleared and the shift register is preloaded again. Therefore, the input signal must remain stable for greater than 1.5625 milliseconds before that transition changes the state of the clock input of the totalizer counter (A1U2-4).

2-48. External Trigger Input Circuits

The External Trigger Input circuit can be configured by the Microprocessor to interrupt on a rising or falling edge of the XT* input (A1J6-2) or to not interrupt on any transitions of the XT* input.

The Microprocessor sets latch output A1U16-19 high for falling edge detection and low for rising edge detection of the XT* input. The Microprocessor can enable the external trigger interrupt by setting port pin A1U4-28 high or disable the interrupt by setting it low. Microprocessor port pin A1U4-28 should only be high if the instrument trigger mode of "ON" has been selected. Resistor A1R20 pulls NAND gate input A1U13 low during power-up to ensure that the external trigger interrupt input (A1U4-9) is high.

When the EXOR gate output (A1U14-3) goes high, and NAND gate input A1U12-13 is high, the output of the NAND gate (A1U12-11) goes low to interrupt the Microprocessor. The Microprocessor can also determine the state of the XT* input by reading the TRIG signal on port pin A1U4-27.

The XT* input is pulled up to +5V dc by A1Z2 and is protected from damage by ESD by A1R58, A1C54, A1Z3, and A1CR15. Capacitor A1C54 helps ensure that the instrument meets EMI/EMC performance requirements.

2-49. A/D Converter PCA

The following paragraphs describe the operation of the circuits on the A/D Converter PCA. The schematic for this pca is located in Section 8.

2-50. Analog Measurement Processor

Refer to Figure 2-3 for an overall picture of the Analog Measurement Processor chip and its peripheral circuits. Table 2-4 describes Analog Measurement Processor chip signal names.

2-16

Page 64
Image 64
Fluke 2635A, 2625A service manual 49. A/D Converter PCA

2635A, 2625A specifications

The Fluke 2625A is a sophisticated data logger designed for a wide range of industrial and laboratory applications. This versatile instrument excels in capturing and monitoring temperature and voltage measurements with precision and reliability. One of its standout features is its ability to record data from various sensors, making it ideal for environments where comprehensive data acquisition is critical.

Equipped with a large LCD display, the Fluke 2625A provides clear and user-friendly access to real-time measurement data. Its graphical interface allows users to visualize trends and monitor readings easily, which significantly enhances usability. The device supports a variety of input types and can connect to thermocouples, RTDs, and other sensors, offering great flexibility for various measurement tasks.

The Fluke 2625A employs advanced measurement technologies that ensure accuracy in data logging. With a high sampling rate, it captures fast-changing signals while maintaining data integrity. The device features a built-in memory that allows for extensive logging, accommodating long-term monitoring tasks without requiring constant oversight. Users can set it to log data at specified intervals, offering customizable solutions for different monitoring needs.

Another significant characteristic of the Fluke 2625A is its robust communication capabilities. The device supports USB and RS-232 interfaces for easy connectivity with computers and other devices, making data transfer and analysis seamless. This feature is especially beneficial for users who need to compile or analyze large sets of data quickly.

Furthermore, the Fluke 2625A is designed with durability in mind. Its rugged construction makes it suitable for challenging environments, including those with high levels of vibration, dust, or moisture. This reliability ensures that the device can withstand the rigors of fieldwork while still delivering accurate measurements.

In summary, the Fluke 2625A is a powerful data logger that combines versatility, accuracy, and reliability. Its user-friendly interface, advanced measurement technologies, and robust design make it an essential tool for professionals in various industries, from manufacturing and HVAC to research and development. Whether monitoring temperature changes in a laboratory or assessing voltage levels in an industrial setting, the Fluke 2625A stands out as a trusted solution for effective data logging and analysis.