Diagnostic Testing and Troubleshooting (2635A) 5A

Digital Kernel Troubleshooting

During instrument power-up, the RESET* and HALT* signals are held low for 140 to 280 milliseconds after the VCC power supply is greater than 4.65 volts dc. Before the Microprocessor can begin execution of the firmware stored in the Flash Memory, the reset circuit must release the RESET* and HALT* signals (A1U2-11 and A1U2-8 respectively) and allow them to go high. Verification of the operation of the RESET* and HALT* signals is best done by using a storage oscilloscope.

After the Microprocessor has begun execution of the instructions stored in Flash Memory (A1U14 and A1U16), the Microprocessor may drive the HALT* signal (A1U1-

91)low if the instructions executed are not correct. Another sign of incorrect instruction execution is the Bus Error signal (BERR*;A1U1-94) going low to indicate that an access to an unused area of memory space was done. To troubleshoot these problems, use an oscilloscope to check the activity of the address, data, and control signals to the Flash Memory devices (A1U14 and A1U16). It may also be useful to check signal continuity by using a DMM with the instrument power off.

To check the Flash Memory control signals, verify that A1U1-128 is going low and is also appearing on pins A1U14-22 and A1U16-22 of the Flash Memory devices. It may be necessary to continually reset (power on) the instrument to check these lines, since the activity probably halts quickly when the instrument software goes awry. Verify that RDU* (A1U11-14 and A1U14-24) goes low when A1U1-128 is low. Verify that RDL* (A1U11-19 and A1U16-24) goes low when A1U1-128 is low. If all this is true, the problem is with the Flash Memory or there is a fault in the address/data lines from the MC68302 Microprocessor.

Verify that the XINIT* output (A1U25-65) goes high after RESET* goes high. Verify that the mode pins and extra chip select input on the FPGA (A1U25) are properly connected to VCC and GND. Pins 6, 29, 54, and 56 must be about 5 volts dc. Pins 52 and 93 must be near 0 volts dc. If the Microprocessor is able to correctly fetch instructions from the Flash Memory, the Microprocessor tries to program the FPGA.

Address decoding for I/O devices like the FPGA is done by A1U11. Verify that the PGA* output (A1U11-12) goes low when the Microprocessor attempts to access the FPGA. Verify the address and I/O* inputs to A1-U11 (pins 2 through 8) according to the decoding shown in the following table.

Output

A<12>

A<11>

A<10>

A<9>

A<8>

A<7>

I/O*

PGA* (A1U11-12)

0

0

0

0

0

0

0

RTC* (A1U11-16)

0

0

0

0

0

1

0

OPTE* (A1U11-12)

0

0

0

0

1

0

0

Programming of the FPGA is initiated by the Microprocessor by driving the XD/P* (A1U25-80) and RESET* (A1U25-78) signals low simultaneously. RESET* is pulsed low by the Microprocessor for approximately 10 microseconds. The Microprocessor then waits for XINIT* (A1U25-65) to go high; if this doesn’t happen, the Watchdog Timer in the Microprocessor will reset the instrument after several seconds by driving POR* (A1U1-117) low. Verify that the Microprocessor waits until XRDY (A1U25-99) is high before writing each byte to the FPGA. (A1U25-88 and A1U25-5 both go low during the write cycle.) Check the XD/P* signal (A1U25-80) at the end of the FPGA programming; if it doesn’t go high, the Microprocessor will repeat the FPGA programming sequence until it works correctly or the Watchdog Timer in the Microprocessor resets the instrument by driving POR* (A1U1-117) low. If FPGA programming is failing, check the D<8> through D<15>, PGA*, WRU*, XINIT*, XRDY, XD/P*, and RESET* signals for activity with an oscilloscope. It may also be necessary to check the continuity of these signals with a DMM when the instrument power is off.

5A-19

Page 219
Image 219
Fluke 2625A, 2635A service manual Output A12 A11 A10

2635A, 2625A specifications

The Fluke 2625A is a sophisticated data logger designed for a wide range of industrial and laboratory applications. This versatile instrument excels in capturing and monitoring temperature and voltage measurements with precision and reliability. One of its standout features is its ability to record data from various sensors, making it ideal for environments where comprehensive data acquisition is critical.

Equipped with a large LCD display, the Fluke 2625A provides clear and user-friendly access to real-time measurement data. Its graphical interface allows users to visualize trends and monitor readings easily, which significantly enhances usability. The device supports a variety of input types and can connect to thermocouples, RTDs, and other sensors, offering great flexibility for various measurement tasks.

The Fluke 2625A employs advanced measurement technologies that ensure accuracy in data logging. With a high sampling rate, it captures fast-changing signals while maintaining data integrity. The device features a built-in memory that allows for extensive logging, accommodating long-term monitoring tasks without requiring constant oversight. Users can set it to log data at specified intervals, offering customizable solutions for different monitoring needs.

Another significant characteristic of the Fluke 2625A is its robust communication capabilities. The device supports USB and RS-232 interfaces for easy connectivity with computers and other devices, making data transfer and analysis seamless. This feature is especially beneficial for users who need to compile or analyze large sets of data quickly.

Furthermore, the Fluke 2625A is designed with durability in mind. Its rugged construction makes it suitable for challenging environments, including those with high levels of vibration, dust, or moisture. This reliability ensures that the device can withstand the rigors of fieldwork while still delivering accurate measurements.

In summary, the Fluke 2625A is a powerful data logger that combines versatility, accuracy, and reliability. Its user-friendly interface, advanced measurement technologies, and robust design make it an essential tool for professionals in various industries, from manufacturing and HVAC to research and development. Whether monitoring temperature changes in a laboratory or assessing voltage levels in an industrial setting, the Fluke 2625A stands out as a trusted solution for effective data logging and analysis.