Theory of Operation (2635A) 2A

Detailed Circuit Description

The SRAM* address decode output (A1U1-127) for the 128 kilobytes of NVRAM goes low for any memory access to A1U20 or A1U24. This signal must go through two NAND gates in A1U26 to the NVRAM chip select inputs (A1U20-22 and A1U24-22). This ensures that when the instrument is powered down and A1U10-7 is driven low, A1U20-22 and A1U24-22 will be driven high so that the contents of the NVRAM cannot be changed and the power dissipated by the NVRAM is minimized. Jumper A1W4 in A<18> is not used in the current instrument; it should be installed only if more NVRAM is needed in a future instrument that needs 512 kilobytes of NVRAM using the same circuit board. A1U24 is connected to the high 8 bits of the data bus, so read accesses are enabled by the Read Upper (RDU*;A1U24-24) signal going low, and write accesses are enabled by the Write Upper (WRU*;A1U24-29) signal going low. A1U20 is connected to the low 8 bits of the data bus, so read accesses are enabled by the Read Lower (RDL*;A1U20-24) signal going low, and write accesses are enabled by the Write Lower (WRL*;A1U20-29) signal going low.

Memory accesses to the Real-Time Clock (A1U12) are enabled by the RTC* address decode output (A1U11-16). This signal must go through two NAND gates in A1U26 to the Real-Time Clock chip select input (A1U12-18). This ensures that when the instrument is powered down and A1U10-7 is driven low, A1U12-18 will be driven high so that the contents of the Real-Time Clock cannot be changed, and the power dissipated by the Real-Time Clock is minimized. A1U12 is connected to the low 8 bits of the data bus, so read accesses are enabled by the Read Lower (RDL*;A1U12-19) signal going low, and write accesses are enabled by the Write Lower (WRL*;A1U12-20) signal going low. When the instrument is powered up, the accuracy of the timebase generated by the internal crystal may be tested by measuring the frequency of the 1-Hz square wave output (A1U12-4). The Real-Time Clock also has an interrupt output (A1U12-3) that is used by the Microprocessor to time the interval between scans when a scan interval is set in the instrument. There should be one interrupt per second from the Real-Time Clock.

2A-37.Serial Communication (Guard Crossing)

The transmission of information from the Microprocessor (A1U1) to the Microcontroller (A3U9) is accomplished via the circuit made up of A1Q10, A1U7, A1R8, A1R16, and A3R8. The transmit output from the Microprocessor (A1U1-54) is buffered by A1Q10, which then switches current through optocoupler LED (A1U7-2). Resistor A1R8 limits the current through the LED.

The phototransistor in A1U7 responds to the light emitted by the LED when A1U1-54 is driven low. (The collector of the phototransistor, A1U7-5, goes low.) The phototransistor collector is pulled up by A3R8 on the A/D Converter PCA. When turning off, the phototransistor base discharges through A1R16. With this arrangement, the rise and fall times of the phototransistor collector signal are nearly symmetrical.

The transmission of data from the Microcontroller (A3U9) to the Microprocessor (A1U1) is accomplished via the circuit made up of A3Q1, A3R7, A1U5, A1R7, and A1R3. The transmit output from the Microcontroller (A3U9-14) is inverted by A3Q1, which drives the optocoupler LED (A1U5-2) through resistor A3R7. The current through the LED is limited by resistor A3R7. The phototransistor in A1U5 responds to the light emitted by the LED when A1U5-2 is driven low. (The collector of the phototransistor, A1U5-4, goes low.) The phototransistor collector (A1U5-5) is pulled up by resistor A1R3. When turning off, the phototransistor base discharges through A1R7. With this arrangement, the rise and fall times of the phototransistor collector signal are nearly symmetrical.

2A-15

Page 97
Image 97
Fluke 2625A, 2635A service manual 2A-37.Serial Communication Guard Crossing

2635A, 2625A specifications

The Fluke 2625A is a sophisticated data logger designed for a wide range of industrial and laboratory applications. This versatile instrument excels in capturing and monitoring temperature and voltage measurements with precision and reliability. One of its standout features is its ability to record data from various sensors, making it ideal for environments where comprehensive data acquisition is critical.

Equipped with a large LCD display, the Fluke 2625A provides clear and user-friendly access to real-time measurement data. Its graphical interface allows users to visualize trends and monitor readings easily, which significantly enhances usability. The device supports a variety of input types and can connect to thermocouples, RTDs, and other sensors, offering great flexibility for various measurement tasks.

The Fluke 2625A employs advanced measurement technologies that ensure accuracy in data logging. With a high sampling rate, it captures fast-changing signals while maintaining data integrity. The device features a built-in memory that allows for extensive logging, accommodating long-term monitoring tasks without requiring constant oversight. Users can set it to log data at specified intervals, offering customizable solutions for different monitoring needs.

Another significant characteristic of the Fluke 2625A is its robust communication capabilities. The device supports USB and RS-232 interfaces for easy connectivity with computers and other devices, making data transfer and analysis seamless. This feature is especially beneficial for users who need to compile or analyze large sets of data quickly.

Furthermore, the Fluke 2625A is designed with durability in mind. Its rugged construction makes it suitable for challenging environments, including those with high levels of vibration, dust, or moisture. This reliability ensures that the device can withstand the rigors of fieldwork while still delivering accurate measurements.

In summary, the Fluke 2625A is a powerful data logger that combines versatility, accuracy, and reliability. Its user-friendly interface, advanced measurement technologies, and robust design make it an essential tool for professionals in various industries, from manufacturing and HVAC to research and development. Whether monitoring temperature changes in a laboratory or assessing voltage levels in an industrial setting, the Fluke 2625A stands out as a trusted solution for effective data logging and analysis.