HYDRA

Service Manual

The Totalizer Debounce circuit in the FPGA (A1U25) allows the Microprocessor to select totalizing of either the input signal or the debounced input signal. The buffered Totalizer Input signal (TOTI*) goes into the FPGA at A1U25-12. Inside the FPGA, the totalizer signal is routed to the Totalizer Output (TOTO, A1U25-8) which then goes to a 16-bit counter in the Microprocessor (A1U1-114; TP20).

The actual debouncing of the input signal is accomplished by A1U25. Counters divide the 12.288-MHz system clock down to 128 kHz for the debouncing circuit. An EXOR gate compares the input signal (TOTI*) and the latched output of the debouncer. If these signals differ, the EXOR gate output goes high, enabling the debouncer. If the input remains stable for 1.75 milliseconds, the totalizer output (TOTO, A1U25-8) changes state. If the input does not remain stable for 1.75 milliseconds, the totalizer output does not change state. For a stable totalizer input of +5V dc, the totalizer output (TOTO, A1U25-8) will be 0.0V dc. For a stable totalizer input of +0.0V dc, the totalizer output (TOTO, A1U25-8) will be +5V dc.

2A-47.External Trigger Input Circuits

The External Trigger Input circuit can be configured by the Microprocessor to interrupt on a rising or falling edge of the XT* input (A1J6-2) or to not interrupt on any transitions of the XT* input. The falling edge of the XT* input is used by the instrument firmware as an indication to start scanning, and the rising edge is used as an indication to stop scanning.

The External Trigger Input is pulled up to +5V dc by A1Z2 and is protected from electrostatic discharge (ESD) damage by A1R58, A1C54, A1Z3, and A1CR15. Capacitor A1C54 helps ensure that the instrument meets EMI/EMC performance requirements.

The input (XTI) is then routed to the FPGA (A1U25), which contains the External Trigger control circuitry. The Microprocessor sets control register bits in the FPGA (A1U25) to control the external trigger circuit. The External Trigger control circuit output (A1U25-9) drives the non-maskable interrupt on the Microprocessor (A1U1-95).

If External Triggering is enabled (see User Manual), the Microprocessor sets FPGA control register bits to allow a low level on the XT* input to cause the External Trigger Interrupt (XTINT*; A1U25-9) to go low. The Microprocessor then changes the FPGA control register bits to allow a high level on the XT* input to cause XTINT* (A1U25-9) to go low. Thus the Microprocessor can detect both rising and falling edges on the XT* input. Normally, the XTINT* output of the FPGA (A1U25-9) should be low only for a few microseconds at any time. If it is held low constantly, the instrument will not be able to operate. Resistor A1R64 pulls the XTINT* output high to ensure that it is high during power-up.

2A-48. A/D Converter PCA

The following paragraphs describe the operation of the circuits on the A/D Converter PCA. The schematic for this pca is located in Section 8.

2A-49.Analog Measurement Processor

Refer to Figure 2A-3 for an overall picture of the Analog Measurement Processor chip and its peripheral circuits. Table 2A-4 describes Analog Measurement Processor chip signal names.

The Analog Measurement Processor (A3U8) is a 68-pin CMOS device that, under control of the A/D Microcontroller (A3U9), performs the following functions:

Input signal routing

Input signal conditioning

Range switching

2A-20

Page 102
Image 102
Fluke 2635A, 2625A 2A-47.External Trigger Input Circuits, 2A-48. A/D Converter PCA, 2A-49.Analog Measurement Processor

2635A, 2625A specifications

The Fluke 2625A is a sophisticated data logger designed for a wide range of industrial and laboratory applications. This versatile instrument excels in capturing and monitoring temperature and voltage measurements with precision and reliability. One of its standout features is its ability to record data from various sensors, making it ideal for environments where comprehensive data acquisition is critical.

Equipped with a large LCD display, the Fluke 2625A provides clear and user-friendly access to real-time measurement data. Its graphical interface allows users to visualize trends and monitor readings easily, which significantly enhances usability. The device supports a variety of input types and can connect to thermocouples, RTDs, and other sensors, offering great flexibility for various measurement tasks.

The Fluke 2625A employs advanced measurement technologies that ensure accuracy in data logging. With a high sampling rate, it captures fast-changing signals while maintaining data integrity. The device features a built-in memory that allows for extensive logging, accommodating long-term monitoring tasks without requiring constant oversight. Users can set it to log data at specified intervals, offering customizable solutions for different monitoring needs.

Another significant characteristic of the Fluke 2625A is its robust communication capabilities. The device supports USB and RS-232 interfaces for easy connectivity with computers and other devices, making data transfer and analysis seamless. This feature is especially beneficial for users who need to compile or analyze large sets of data quickly.

Furthermore, the Fluke 2625A is designed with durability in mind. Its rugged construction makes it suitable for challenging environments, including those with high levels of vibration, dust, or moisture. This reliability ensures that the device can withstand the rigors of fieldwork while still delivering accurate measurements.

In summary, the Fluke 2625A is a powerful data logger that combines versatility, accuracy, and reliability. Its user-friendly interface, advanced measurement technologies, and robust design make it an essential tool for professionals in various industries, from manufacturing and HVAC to research and development. Whether monitoring temperature changes in a laboratory or assessing voltage levels in an industrial setting, the Fluke 2625A stands out as a trusted solution for effective data logging and analysis.