Table 5A-3. Power Supply Troubleshooting Guide (2635A)

Symptom

Line fuse blows.

Fault

-Shorted A1CR2 or A1CR3.

-Shorted A1CR10.

-Shorted A1C7.

-Shorted A1C26.

Supply voltage for A1U23 and A1U22 is greater than 7V (7 to 30V).

VCC (5.1V) supply is at the raw supply level (7.5 to 35V dc).

VCC (5.1V) supply shows excessive ripple (about 1V p-p).

VCC is below approximately 4.5V. Duty cycle of 5V switcher supply is very low (ON time near 0.1).

VCC is about 1.5V. 5V switcher supply is in current limit.

VCC is below approximately 1V. 5V switcher supply is in current limit, with very low duty cycle (ON time near 0.1).

Input-to-output short of A1U19. This fault may have caused damage to A1Q7 and A1Q8.

Shorted switch transistor in A1U9 (A1U9-5 to 7). Open A1C26 can cause switch transistor to short.

A1C14 open.

Drain-to-source short of A1Q7 or A1Q8.

Shorted A1CR5 or A1CR6.

Shorted A1C14.

VCC is below approximately 4.5V. 5V switcher supply is in current limit, with very low duty cycle (ON time near 0.1).

-Q or Q* output of A1U22 stuck high.

-A1U23 pin 8 output stuck high or low.

-Shorted A1CR7

-Shorted A1CR9 (either diode), pins 1-3 or 2-3.

-Shorted A1C30. A1CR13 may also be damaged.

-Shorted A1C31. A1CR13 may also be damaged.

-Shorted A1C12.

-Shorted A1C13.

-Shorted A1CR8 (either diode), pins 1-3 or 2-3.

VLOAD (-30V dc) Inverter Supply is at -36V.

VLOAD (-30V dc) Inverter Supply is OFF.

VLOAD (-30V dc) Inverter Supply ripple.

VDD (5.3V dc) supply at approximately 9.2V.

VSS (-5.4V dc) supply at approximately -9.2V.

VDDR (5.6V dc) supply at approximately 10V.

VDDR supply has 4-to-5 volt spikes when the A/D relays are switched (set or reset).

VEE (-5V dc) supply is low (near zero).

A1CR13, Diode 1-3 shorted or open.

VEE supply is high (near -9V).

A1U18 input has large square wave component.

Q output of A1U22 stuck low.

Q* output of A1U22 stuck low.

-Open A1CR8 (either diode).

-Open A1CR9 (either diode).

Emitter-to-collector short of A1Q2.

Emitter-to-collector short of A1Q5.

Input-to-output short of A1U6.

Open A1C12.

-Open A1C30.

-A1CR13 open.

A1C30 may be shorted.

Input-to-Output short of A1U18.

Open A1C31.

5A-12

Page 212
Image 212
Fluke 2625A service manual Power Supply Troubleshooting Guide 2635A Symptom, Fault

2635A, 2625A specifications

The Fluke 2625A is a sophisticated data logger designed for a wide range of industrial and laboratory applications. This versatile instrument excels in capturing and monitoring temperature and voltage measurements with precision and reliability. One of its standout features is its ability to record data from various sensors, making it ideal for environments where comprehensive data acquisition is critical.

Equipped with a large LCD display, the Fluke 2625A provides clear and user-friendly access to real-time measurement data. Its graphical interface allows users to visualize trends and monitor readings easily, which significantly enhances usability. The device supports a variety of input types and can connect to thermocouples, RTDs, and other sensors, offering great flexibility for various measurement tasks.

The Fluke 2625A employs advanced measurement technologies that ensure accuracy in data logging. With a high sampling rate, it captures fast-changing signals while maintaining data integrity. The device features a built-in memory that allows for extensive logging, accommodating long-term monitoring tasks without requiring constant oversight. Users can set it to log data at specified intervals, offering customizable solutions for different monitoring needs.

Another significant characteristic of the Fluke 2625A is its robust communication capabilities. The device supports USB and RS-232 interfaces for easy connectivity with computers and other devices, making data transfer and analysis seamless. This feature is especially beneficial for users who need to compile or analyze large sets of data quickly.

Furthermore, the Fluke 2625A is designed with durability in mind. Its rugged construction makes it suitable for challenging environments, including those with high levels of vibration, dust, or moisture. This reliability ensures that the device can withstand the rigors of fieldwork while still delivering accurate measurements.

In summary, the Fluke 2625A is a powerful data logger that combines versatility, accuracy, and reliability. Its user-friendly interface, advanced measurement technologies, and robust design make it an essential tool for professionals in various industries, from manufacturing and HVAC to research and development. Whether monitoring temperature changes in a laboratory or assessing voltage levels in an industrial setting, the Fluke 2625A stands out as a trusted solution for effective data logging and analysis.