Diagnostic Testing and Troubleshooting (2635A) 5A

Calibration Failures

5A-23. Retrieving Calibration Constants

If a calibration error is suspected, the stored constant can be retrieved and verified over the computer interface. Acceptable calibration constants for each function and range are listed in Table 5A-9. Retrieve the constant with the following command:

CAL_CONST? xx

(where xx denotes the calibration constant number)

The entire calibration of the Hydra Databucket can be retrieved from the instrument in tabular form by using the following command:

EEM_LIST?

The instrument response from this command shows the currently used FLASH memory block and page numbers, and each calibration constant in hexadecimal and floating point with a description of what function and range each calibration constant is used on. The following is a sample of a typical EEM_LIST? response:

Page 10 of Parameter Block 1 is currently in use.

Register Hex Value F.P. Value Description

---------------------------------------------------------------

0-1

0x3F8A847A1.

0821679100

Millivolt DC Gain

2-3

0x3F8A5D751.

08097711

Volt DC Gain

4-5

0x3F8A63571.

0811566300

Millivolt DC Gain

6-7

0x3F8A5BC01.

08092503

Volt DC Gain

8-9

0x3F8B4F571.

088358830

Volt DC Gain

10-11

0x3F8A7C8A1.

0819256150/300

Volt DC Gain

12-13

0xB8F20A00

-0.0001154300

Millivolt AC Offset

14-15

0x3F8A51191.

0805999300

Millivolt AC Gain

16-17

0xBA974800

-0.00115423

Volt AC Offset

18-19

0x3F8A7F291.

08200563

Volt AC Gain

20-21

0xBC3A6C00

-0.011378330

Volt AC Offset

22-23

0x3F8A863A1.

082221330

Volt AC Gain

24-25

0xBDE81400

-0.1133194150/300

Volt AC Offset

26-27

0x3F8AC8861.

0842445150/300

Volt AC Gain

28-29

0x3F808EAB1.

0043539

300 Ohm Gain

30-31

0x3F8097591.

0046188

3 Kilo Ohm Gain

32-33

0x3F80FD781.

0077353

30 Kilo Ohm Gain

34-35

0x3F7FF46C0.

9998233

300 Kilo Ohm Gain

36-37

0x3F7FFDD00.

9999666

3 Mega Ohm Gain

38-39

0x3F7FFDD00.

9999666

10/30 Mega Ohm Gain

40-41

0x3F80073D1.

0002209

Frequency Gain

42

0x000F

 

Calibration Status

43-44

0xAAAAAAAA

 

Product Serial Number

45

0xFFFF

 

Unused EEM Register #1

46

0xFFFF

 

Unused EEM Register #2

47

0xFFFF

 

Unused EEM Register #3

48

0x62F4

 

CRC of EEM Data (0x62F4)

5A-24. Replacing the Flash Memory (A1U14 and A1U16)

The FLASH Memory provides nonvolatile storage for the instrument serial number, the instrument firmware, and all calibration information. If the "boot" firmware in FLASH memory has been determined to be faulty, A1U14 and A1U16 must both be replaced.

Many other problems may be corrected by loading new instrument firmware in the instrument (see the section entitled "Updating the 2635A Instrument Firmware" in Section 4 of this manual), or recalibrating the instrument.

If the FLASH Memories must be replaced during repair, the instrument must be recalibrated. The new FLASH Memory can be programmed with the 7-digit serial number found on the rear panel of the instrument or any 7-digit identifier of your choosing. Note that the serial number is not programmed prior to shipment from the factory.

5A-29

Page 229
Image 229
Fluke 2625A, 2635A service manual 5A-23. Retrieving Calibration Constants, 5A-24. Replacing the Flash Memory A1U14 and A1U16

2635A, 2625A specifications

The Fluke 2625A is a sophisticated data logger designed for a wide range of industrial and laboratory applications. This versatile instrument excels in capturing and monitoring temperature and voltage measurements with precision and reliability. One of its standout features is its ability to record data from various sensors, making it ideal for environments where comprehensive data acquisition is critical.

Equipped with a large LCD display, the Fluke 2625A provides clear and user-friendly access to real-time measurement data. Its graphical interface allows users to visualize trends and monitor readings easily, which significantly enhances usability. The device supports a variety of input types and can connect to thermocouples, RTDs, and other sensors, offering great flexibility for various measurement tasks.

The Fluke 2625A employs advanced measurement technologies that ensure accuracy in data logging. With a high sampling rate, it captures fast-changing signals while maintaining data integrity. The device features a built-in memory that allows for extensive logging, accommodating long-term monitoring tasks without requiring constant oversight. Users can set it to log data at specified intervals, offering customizable solutions for different monitoring needs.

Another significant characteristic of the Fluke 2625A is its robust communication capabilities. The device supports USB and RS-232 interfaces for easy connectivity with computers and other devices, making data transfer and analysis seamless. This feature is especially beneficial for users who need to compile or analyze large sets of data quickly.

Furthermore, the Fluke 2625A is designed with durability in mind. Its rugged construction makes it suitable for challenging environments, including those with high levels of vibration, dust, or moisture. This reliability ensures that the device can withstand the rigors of fieldwork while still delivering accurate measurements.

In summary, the Fluke 2625A is a powerful data logger that combines versatility, accuracy, and reliability. Its user-friendly interface, advanced measurement technologies, and robust design make it an essential tool for professionals in various industries, from manufacturing and HVAC to research and development. Whether monitoring temperature changes in a laboratory or assessing voltage levels in an industrial setting, the Fluke 2625A stands out as a trusted solution for effective data logging and analysis.