Theory of Operation (2635A) 2A

Detailed Circuit Description

The output voltage of the switcher supply is controlled by varying the duty cycle (ON time) of the switching transistor in the controller/switch device A1U9. A1U9 contains the supply reference, oscillator, switch transistor, pulse-width modulator comparator, switch drive circuit, current-limit comparator, current-limit reference, and thermal limit. Dual inductor A1T2 regulates the current that flows from the raw supply to the load as the switching transistor in A1U9 is turned on and off. Complementary switchA1CR10 conducts when the switching transistor is off.

The pulse-width modulator comparator in A1U9 compares the output to the reference and sets the ON-time/OFF-time ratio to regulate the output to 5.1V dc. A1C26 is the input filter capacitor, and A1C14 is the output filter capacitor. Proper inductor and capacitor values set the filter frequency response to ensure best overall system stability. Circuitry consisting of A1R26, A1C21, and A1C18 ensure that the switcher supply remains stable and operating in the continuous mode. Resistors A1R30 and A1R31 set the output voltage to within 5% of 5.1V.Capacitor A1C21 sets the operating frequency of the switcher at approximately 100 kHz.

Resistors A1R30 and A1R31 form a voltage divider that operates in conjunction with amplifier A1U28, which is configured as a voltage follower.A1U28-5 samples the 5.1V dc output, while A1U28-6 is the voltage divider input.The effect is to maintain the junction of R30 and R31 at 5.1V dc, resulting in an A1U28-7 output level of 6.34V dc, or 1.24V dc above the output.This feedback voltage is applied to A1U9-2, which A1U9 interprets as 1.24V dc because A1U9-3 (ground) is connected to the 5.1V dc output.

A1U9 maintains the feedback and reference voltages at 1.24V dc and thus regulates the 5.1V dc source.

2A-27.Inverter

The inverter supply uses a two transistor driven push-pull configuration. The center tap of transformer A1T1 primary is connected to the 5.0V dc VCC supply, and each side is alternately connected to common through transistors A1Q7 and A1Q8. A1R38 may be removed to disable the inverter supply for troubleshooting purposes. A1Q7 and A1Q8 are driven by the outputs of D flip-flop A1U22. Resistors A1R34 and A1R28, and diodes A1CR11 and A1CR12 shape the input drive signals to properly drive the gate of the transistors. D flip-flop A1U22 is wired as a divide-by-two counter driven by a 110-kHz square wave. The 110-kHz square wave is generated by hex inverter A1U23, which is connected as an oscillator with a frequency determined by the values of resistors A1R40 and A1R47 and capacitor A1C35. The resulting ac voltage produced across the secondary of A1T1 is rectified to provide the input to the inverter inguard and outguard supplies.

2A-28.Inverter Outguard Supply

The inverter outguard supply provides three outputs: 5.4V ac, -30V dc, and -5V dc. These voltages are required by the display and RS-232 drivers and receiver. The 5.4V ac supply comes off the secondary windings (pins 6 and 7) on transformer T1, and it is biased at -24V dc with zener diode A1VR3 and resistor A1R22. Dual diodes A1CR8 and A1CR9 and capacitor A1C17 are for the -30V dc supply. Capacitors A1C30 and A1C31, and dual diodes A1CR13 form a voltage doubler circuit that generates -12 volts. Three- terminal regulator A1U18 then regulates this voltage down to -5V for the RS-232 circuit. Capacitor A1C32 is needed for transient response performance of the three-terminal regulator.

2A-9

Page 91
Image 91
Fluke 2625A, 2635A service manual 2A-27.Inverter

2635A, 2625A specifications

The Fluke 2625A is a sophisticated data logger designed for a wide range of industrial and laboratory applications. This versatile instrument excels in capturing and monitoring temperature and voltage measurements with precision and reliability. One of its standout features is its ability to record data from various sensors, making it ideal for environments where comprehensive data acquisition is critical.

Equipped with a large LCD display, the Fluke 2625A provides clear and user-friendly access to real-time measurement data. Its graphical interface allows users to visualize trends and monitor readings easily, which significantly enhances usability. The device supports a variety of input types and can connect to thermocouples, RTDs, and other sensors, offering great flexibility for various measurement tasks.

The Fluke 2625A employs advanced measurement technologies that ensure accuracy in data logging. With a high sampling rate, it captures fast-changing signals while maintaining data integrity. The device features a built-in memory that allows for extensive logging, accommodating long-term monitoring tasks without requiring constant oversight. Users can set it to log data at specified intervals, offering customizable solutions for different monitoring needs.

Another significant characteristic of the Fluke 2625A is its robust communication capabilities. The device supports USB and RS-232 interfaces for easy connectivity with computers and other devices, making data transfer and analysis seamless. This feature is especially beneficial for users who need to compile or analyze large sets of data quickly.

Furthermore, the Fluke 2625A is designed with durability in mind. Its rugged construction makes it suitable for challenging environments, including those with high levels of vibration, dust, or moisture. This reliability ensures that the device can withstand the rigors of fieldwork while still delivering accurate measurements.

In summary, the Fluke 2625A is a powerful data logger that combines versatility, accuracy, and reliability. Its user-friendly interface, advanced measurement technologies, and robust design make it an essential tool for professionals in various industries, from manufacturing and HVAC to research and development. Whether monitoring temperature changes in a laboratory or assessing voltage levels in an industrial setting, the Fluke 2625A stands out as a trusted solution for effective data logging and analysis.