HYDRA

Service Manual

To provide accuracy at full range, calibration is not recommended below one-third of full range (10000 counts).

Command

Cal x

Table 4-8. Calibration Mode Computer Interface Commands

Description

Start calibration of a new function.

x

Function to calibrate

 

 

1VDC

2VAC

3ohms

4Frequency

CAL_CLR

CAL_CONST? xx

CAL_REF?

CAL_REF < value>

CAL_STEP?

EEREG? xx

Reset all calibration constants to nominal values, clearing present calibration.

Return the value of the calibration constant indicated by xx.

Return the present calibration reference.

Calibrate to <value>, rather than the default calibration reference value.

Calibrate and return the calibrated value of the input.

Return the contents of the specified EEPROM register (xx).

The following additional computer interface commands can be used in calibration mode. Use of any other command results in an execution error. Refer to Section 4 of the Hydra Users Manual for complete information about these computer interface commands.

*CLS *ESE *ESE?*ESR?*IDN?*OPC *OPC?*RST *SRE *SRE?*STB?*TRG *WAI IEE IEE?IER?LOCS LWLS REMS RWLS

4-21. Using Hydra Starter Calibration Software

This procedure uses the Hydra Starter Package (with Calibration Software) for closed- case calibration. This software runs on an IBM PC or equivalent using the RS-232 interface. It consists of the following three files:

An executable file (CAL.EXE)

A help text file (CAL.HLP)

A configuration initialization file (CAL.INI)

4-22. Setup Procedure Using Starter

Use the following procedure to set up Hydra and the PC prior to using the calibration feature of the Hydra Starter Package:

1.Connect a pair of test leads to the channel 1 H (high) and L (low) terminals on the Input Module. Connect a second pair of test leads to channel 11. (This second pair is used only for 4-wire resistance calibration.) Install the Input Module into Hydra.

2.Using the RS40 Terminal Interface Cable, connect the PC COM port to the Hydra RS-232 port. Use an RS40 and an RS41 cable in series to connect to the COM port on an IBM PC/AT.

4-20

Page 160
Image 160
Fluke 2635A, 2625A service manual Using Hydra Starter Calibration Software, Setup Procedure Using Starter, Command

2635A, 2625A specifications

The Fluke 2625A is a sophisticated data logger designed for a wide range of industrial and laboratory applications. This versatile instrument excels in capturing and monitoring temperature and voltage measurements with precision and reliability. One of its standout features is its ability to record data from various sensors, making it ideal for environments where comprehensive data acquisition is critical.

Equipped with a large LCD display, the Fluke 2625A provides clear and user-friendly access to real-time measurement data. Its graphical interface allows users to visualize trends and monitor readings easily, which significantly enhances usability. The device supports a variety of input types and can connect to thermocouples, RTDs, and other sensors, offering great flexibility for various measurement tasks.

The Fluke 2625A employs advanced measurement technologies that ensure accuracy in data logging. With a high sampling rate, it captures fast-changing signals while maintaining data integrity. The device features a built-in memory that allows for extensive logging, accommodating long-term monitoring tasks without requiring constant oversight. Users can set it to log data at specified intervals, offering customizable solutions for different monitoring needs.

Another significant characteristic of the Fluke 2625A is its robust communication capabilities. The device supports USB and RS-232 interfaces for easy connectivity with computers and other devices, making data transfer and analysis seamless. This feature is especially beneficial for users who need to compile or analyze large sets of data quickly.

Furthermore, the Fluke 2625A is designed with durability in mind. Its rugged construction makes it suitable for challenging environments, including those with high levels of vibration, dust, or moisture. This reliability ensures that the device can withstand the rigors of fieldwork while still delivering accurate measurements.

In summary, the Fluke 2625A is a powerful data logger that combines versatility, accuracy, and reliability. Its user-friendly interface, advanced measurement technologies, and robust design make it an essential tool for professionals in various industries, from manufacturing and HVAC to research and development. Whether monitoring temperature changes in a laboratory or assessing voltage levels in an industrial setting, the Fluke 2625A stands out as a trusted solution for effective data logging and analysis.