5-10. Analog Troubleshooting

Warning

To avoid electric shock, disconnect all channelinputs from the instrument before performing anytroubleshooting operations.

Refer to Figure 5-4 and Figure 5-5 for test point locations on the A/D Converter PCA.

First, check for analog-related errors displayed at power up. An ’Error 9’ means that the Main Microprocessor A1U4 is not able to communicate with the A/D Microcontroller A3U9. ’Error A’ and ’Error b’ mean that a failure has occurred in the internal memory of the A/D Microcontroller A3U9. ’Error C’ means that the Analog Measurement Processor A3U8 is not functioning properly.

Check the inguard power supplies on the Main PCA with and without the A/D Converter PCA connected. The inguard supplies must be measured with respect to COM testpoint A1TP30.

Power Supply

Test Location

Acceptable Range

VDD

A1TP31

+5.00 to 5.70V dc

VSS

A1TP32

-5.10 to -5.75V dc

VDDR

A1C6

5.30 to 5.95V dc

Check the inguard supply voltages on the A/D Converter PCA with respect to A3TP9. The following table lists the components nearest the power supply test points.

Power Supply

Test Location

Acceptable Range

VDD

A3C8

5.00 to 5.70V dc

VSS

A3C9

-5.10 to -5.75V dc

VDDR

A3C19

5.30 to 5.95V dc

+VAC

A3CR1

4.7 to 5.7V dc

-VAC

A3C26

-4.8 to -5.7V dc

5-12

Page 182
Image 182
Fluke 2635A, 2625A service manual Analog Troubleshooting, Power Supply Test Location Acceptable Range

2635A, 2625A specifications

The Fluke 2625A is a sophisticated data logger designed for a wide range of industrial and laboratory applications. This versatile instrument excels in capturing and monitoring temperature and voltage measurements with precision and reliability. One of its standout features is its ability to record data from various sensors, making it ideal for environments where comprehensive data acquisition is critical.

Equipped with a large LCD display, the Fluke 2625A provides clear and user-friendly access to real-time measurement data. Its graphical interface allows users to visualize trends and monitor readings easily, which significantly enhances usability. The device supports a variety of input types and can connect to thermocouples, RTDs, and other sensors, offering great flexibility for various measurement tasks.

The Fluke 2625A employs advanced measurement technologies that ensure accuracy in data logging. With a high sampling rate, it captures fast-changing signals while maintaining data integrity. The device features a built-in memory that allows for extensive logging, accommodating long-term monitoring tasks without requiring constant oversight. Users can set it to log data at specified intervals, offering customizable solutions for different monitoring needs.

Another significant characteristic of the Fluke 2625A is its robust communication capabilities. The device supports USB and RS-232 interfaces for easy connectivity with computers and other devices, making data transfer and analysis seamless. This feature is especially beneficial for users who need to compile or analyze large sets of data quickly.

Furthermore, the Fluke 2625A is designed with durability in mind. Its rugged construction makes it suitable for challenging environments, including those with high levels of vibration, dust, or moisture. This reliability ensures that the device can withstand the rigors of fieldwork while still delivering accurate measurements.

In summary, the Fluke 2625A is a powerful data logger that combines versatility, accuracy, and reliability. Its user-friendly interface, advanced measurement technologies, and robust design make it an essential tool for professionals in various industries, from manufacturing and HVAC to research and development. Whether monitoring temperature changes in a laboratory or assessing voltage levels in an industrial setting, the Fluke 2625A stands out as a trusted solution for effective data logging and analysis.