Diagnostic Testing and Troubleshooting (2635A) 5A

Calibration Failures

2.Wait a moment for the instrument to beep, then release SHIFT. The entire display will now stay on until you are ready to deactivate it.

3.At the end of the activation period, press any button on the front panel; the instrument resumes the mode in effect prior to the power interruption (Active or Inactive.)

5A-20. Calibration Failures

5A-21. Introduction

Calibration of Hydra through the computer interface is described in Section 4 of this manual. Generally, a calibration failure is indicated by a Device Dependent Error and a "!>" prompt after a CAL_STEP? command if the RS-232 interface is being used. This occurs if the analog input varies from what the instrument expects to see by more than ±5% or ±15%, depending on the calibration step.

Before suspecting a fault with Hydra, verify that the calibration is being conducted properly.

Check the connections between the source and the instrument. Are allthe connections in place?

Check the output of the calibration source. Does it equal the valuecalled for by this calibration step?

Check the calibration source. Is it in operate mode? Has it revertedto standby?

If a calibration step has failed, Hydra remains on that step so that the output from the calibration source may be corrected or the calibration reference value (CAL_REF) being used by Hydra may be changed if it was improperly entered. The calibration step may be repeated by sending the CAL_STEP? command to Hydra again.

Calibration of Hydra utilizes a simple "calibration by function" approach. If you suspect calibration errors, but the instrument does not exhibit the symptoms mentioned above, verify that you are observing the following calibration rules:

Independent calibration of any function results in the storage ofcalibration constants for that function only.

Once calibration is begun, all steps for that function must becompleted before the calibration constants are stored. If all stepsare not completed and the procedure is terminated, no constants forthat function are stored; only calibration constants for previouslycompleted functions are stored.

5A-22. Calibration-Related Components

If the calibration setup is correct, a faulty component within Hydra may be causing the failure. Each measurement function depends on a combination of components in and around the Analog Measurement Processor (A3U8).

RMS Converter

A3U6

AC Buffer

A3U7

Zener Reference

A3VR1

Divider Network (DC/Ohms)

A3Z4

Integrate Resistors, Reference Divider

A3Z2

AC Divider Network

A3Z3

RMS Converter Network

A3Z1

5A-27

Page 227
Image 227
Fluke 2625A, 2635A service manual 5A-20. Calibration Failures, 5A-21. Introduction, 5A-22. Calibration-Related Components

2635A, 2625A specifications

The Fluke 2625A is a sophisticated data logger designed for a wide range of industrial and laboratory applications. This versatile instrument excels in capturing and monitoring temperature and voltage measurements with precision and reliability. One of its standout features is its ability to record data from various sensors, making it ideal for environments where comprehensive data acquisition is critical.

Equipped with a large LCD display, the Fluke 2625A provides clear and user-friendly access to real-time measurement data. Its graphical interface allows users to visualize trends and monitor readings easily, which significantly enhances usability. The device supports a variety of input types and can connect to thermocouples, RTDs, and other sensors, offering great flexibility for various measurement tasks.

The Fluke 2625A employs advanced measurement technologies that ensure accuracy in data logging. With a high sampling rate, it captures fast-changing signals while maintaining data integrity. The device features a built-in memory that allows for extensive logging, accommodating long-term monitoring tasks without requiring constant oversight. Users can set it to log data at specified intervals, offering customizable solutions for different monitoring needs.

Another significant characteristic of the Fluke 2625A is its robust communication capabilities. The device supports USB and RS-232 interfaces for easy connectivity with computers and other devices, making data transfer and analysis seamless. This feature is especially beneficial for users who need to compile or analyze large sets of data quickly.

Furthermore, the Fluke 2625A is designed with durability in mind. Its rugged construction makes it suitable for challenging environments, including those with high levels of vibration, dust, or moisture. This reliability ensures that the device can withstand the rigors of fieldwork while still delivering accurate measurements.

In summary, the Fluke 2625A is a powerful data logger that combines versatility, accuracy, and reliability. Its user-friendly interface, advanced measurement technologies, and robust design make it an essential tool for professionals in various industries, from manufacturing and HVAC to research and development. Whether monitoring temperature changes in a laboratory or assessing voltage levels in an industrial setting, the Fluke 2625A stands out as a trusted solution for effective data logging and analysis.