Diagnostic Testing and Troubleshooting (2635A) 5A

Analog Troubleshooting

Table 5A-4. DC Volts HI Troubleshooting (2635A)

Checkpoint

A3R11 HI

A3U8 pin 23

A3U8 pin 58

Signal Description

Input

Input

Input, DC filter output

Possible Fault

A3K1 through A3K14, A3U4, A3U5, A3U11, A3U12, A3L1, A3L2, A3L3

A3R11, A3K17, A3R42, A3C32

A3U8, A3Q2

5A-12. AC Volts Troubleshooting

Setup the instrument to measure a channel on the 300 mV ac range, and apply a signal to that channel. Then trace this HI signal (referenced to the input channel LO terminal) as described in Table 5A-5.

If the input HI path traces out properly, remove the input from the channel, and trace continuity through the LO path. Check among A3L4 through A3L24, A3K1 through A3K14, A3R43, A3R34, A3K16, and A3U8 pin 13.

Table 5A-5. AC Volts HI Troubleshooting (2635A)

Checkpoint

A3R11 HI

A3Z3 pin 1

A3U6 pin 13

Signal Description

Input

Input

Amplified (X 2.5) input

Possible Fault

A3K1 through A3K14, A3U4, A3U5, A3U11, A3U12 A3L1, A3L2, A3L3

A3R11, A3C31, A3K15

A3U7, A3Z3, A3Q3 through A3Q9, A3C15, A3C16, A3R24, A3A25, A3R26, A3R27, A3R28, A3C23, A3U6, A3Q13, A3U8

A3Z1 pin 2

A3U8 pin 61

DC equivalent of original input

A3Z1, A3U8, A3R20, A3C6, A3C7, A3C10, A3R16,

 

A3R17

 

 

DC equivalent of original input

A3Z1, A3U8, A3R20, A3C6, A3C7, A3C10, A3R16,

 

A3R17

 

 

5A-13. Ohms Troubleshooting

Setup a channel with an open input for the desired ohms range and place the instrument in monitor mode on that channel. Use a meter with high input impedance to measure the open-circuit voltage at the channel input for the ohms range as listed in Table 5A-6. If a high input impedance meter is not available, only the 30-kΩand lower ranges can be checked.

If the proper voltage is not measured, setup a channel on the 300Ω range (open input), and have the instrument monitor that channel. Check for 3V dc with respect to A3TP9, and work through the HI SOURCE and HI SENSE paths as described in Table 5A-7.

If the HI path works correctly, trace continuity through the LO path. Check among A3L4 through A3L24, A3K1 through A3K14, A3R35, A3U8 pin 11, A3R43, A3K16, A3R34, and A3U8 pin 13.

5A-17

Page 217
Image 217
Fluke 2625A 5A-12. AC Volts Troubleshooting, 5A-13. Ohms Troubleshooting, DC Volts HI Troubleshooting 2635A Checkpoint

2635A, 2625A specifications

The Fluke 2625A is a sophisticated data logger designed for a wide range of industrial and laboratory applications. This versatile instrument excels in capturing and monitoring temperature and voltage measurements with precision and reliability. One of its standout features is its ability to record data from various sensors, making it ideal for environments where comprehensive data acquisition is critical.

Equipped with a large LCD display, the Fluke 2625A provides clear and user-friendly access to real-time measurement data. Its graphical interface allows users to visualize trends and monitor readings easily, which significantly enhances usability. The device supports a variety of input types and can connect to thermocouples, RTDs, and other sensors, offering great flexibility for various measurement tasks.

The Fluke 2625A employs advanced measurement technologies that ensure accuracy in data logging. With a high sampling rate, it captures fast-changing signals while maintaining data integrity. The device features a built-in memory that allows for extensive logging, accommodating long-term monitoring tasks without requiring constant oversight. Users can set it to log data at specified intervals, offering customizable solutions for different monitoring needs.

Another significant characteristic of the Fluke 2625A is its robust communication capabilities. The device supports USB and RS-232 interfaces for easy connectivity with computers and other devices, making data transfer and analysis seamless. This feature is especially beneficial for users who need to compile or analyze large sets of data quickly.

Furthermore, the Fluke 2625A is designed with durability in mind. Its rugged construction makes it suitable for challenging environments, including those with high levels of vibration, dust, or moisture. This reliability ensures that the device can withstand the rigors of fieldwork while still delivering accurate measurements.

In summary, the Fluke 2625A is a powerful data logger that combines versatility, accuracy, and reliability. Its user-friendly interface, advanced measurement technologies, and robust design make it an essential tool for professionals in various industries, from manufacturing and HVAC to research and development. Whether monitoring temperature changes in a laboratory or assessing voltage levels in an industrial setting, the Fluke 2625A stands out as a trusted solution for effective data logging and analysis.