Fluke 2635A, 2625A service manual A Specifications, DC Voltage Inputs Resolution Range Slow Fast

Models: 2635A 2625A

1 310
Download 310 pages 12.8 Kb
Page 36
Image 36

HYDRA

Service Manual

Table 1-4. 2635A Specifications

The instrument specifications presented here are applicable within the conditions listed in the Environmental portion of this specification.

The specifications state total instrument accuracy following calibration, including:

A/D errors

Linearization conformity

Initial calibration errors

Isothermality errors

Relay thermal emf’s

Reference junction conformity

Temperature coefficients

Humidity errors

Sensor inaccuracies are not included in the accuracy figures.

Accuracies at Temperatures Other Than Specified

To determine typical accuracies at temperatures intermediate to those listed in the specification tables, linearly interpolate between the applicable 0°C to 60°C and 18°C to 28°C accuracy specifications.

Response Times

Refer to Typical Scanning Rate and Maximum Autoranging Time later in this table.

DC Voltage Inputs

Resolution

Range

Slow

Fast

 

 

90 mV*

 

 

 

1 μV

 

 

 

10 μV

 

 

300 mV

 

 

 

10 μV

 

 

 

0.1 mV

 

 

 

 

 

 

 

 

 

 

3V

 

 

 

0.1 mV

 

 

 

1 mV

 

 

 

 

 

 

 

 

 

 

30V

 

 

 

1 mV

 

 

 

10 mV

 

 

150/300V

 

 

 

10 mV

 

 

 

0.1V

 

 

 

 

 

 

 

 

 

 

900V* **

 

 

 

10 μV

 

 

 

0.1 mV

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Accuracy ±(% ±V)

 

 

 

 

Range

 

18°C to 28°C

 

 

 

 

0°C to 60°C

 

 

90 Days, Slow

1 Year, Slow

 

1 Year, Fast

1 Year, Slow

1 Year, Fast

90 mV*

0.29% + 7μV

0.034%

+ 7 μV

0.054%

+ 20 μV

300 mV

0.026% + 20 μV

0.031%

+ 20 μV

0.047%

+ 0.2 mV

3V

0.028%

+ 0.2 mV

0.033%

+ 0.2 mV

0.050%

+ 2 mV

30V

0.024%

+ 2 mV

0.029%

+ 2 mV

0.046%

+ 20 mV

150/300V

0.023%

+ 20 mV

0.028%

+ 20 mV

0.045%

+ 0.2V

900 mV

0.026%

+ 20 μV

0.031%

+ 21 μV

0.047%

+ 0.2 mV

*Not used in Autoranging.

**Computer interface only (see FUNC command).

0.074% + 7 μV

0.070% + 20 μV

0.072% + 0.2 mV

0.090% + 2 mV

0.090% + 20 mV

0.070% + 20 μV

0.094% + 20 μV

0.087% + 0.2 mV

0.089% + 2 mV

0.107% + 20 mV

0.107% + 0.2V

0.087% +0.2 mV

1-20

Page 36
Image 36
Fluke 2635A, 2625A service manual A Specifications, DC Voltage Inputs Resolution Range Slow Fast