Performance Testing and Calibration

4

Updating 2635A Data Bucket Embedded Instrument Firmware

Table 4-11. 4-Wire Ohms Calibration (Fixed Resistor)

Command

CAL 3

CAL_REF?

CAL_STEP?

Response

=>

+290.00E+0

Action

Puts Hydra in OHMS Calibration.

You source 290Ω from the decade resistance source or fixed resistor.

Hydra computes calibration constant 15 and returns the calibrated reading. (For software versions lower than 5.4, this step computes calibration constant 13.)

Note

If the input is incorrect, the "!>" response signifies that a Device Dependent Error has been generated. The calibration step could not be executed.Verify that the input to Hydra channel 1 is the correct value. If the input is correct, Hydra may require repair.

CAL_REF?

CAL_STEP?

CAL_REF?

CAL_STEP?

CAL_REF?

CAL_STEP?

CAL_REF?

CAL_STEP?

+2.9000E+3

+29.000E+3

+290.00E+3

+2.9000E+6

You source 2900Ω.

Hydra computes calibration constant 16 and returns the calibrated reading. (For software versions lower than 5.4, this step computes calibration constant 14.)

You source 29000Ω.

Hydra computes calibration constant 17 and returns the calibrated reading. (For software versions lower than 5.4, this step computes calibration constant 15.)

You source 290000Ω.

Hydra computes calibration constant 18 and returns the calibrated reading. (For software versions lower than 5.4, this step computes calibration constant 16.)

You source 2900000Ω.

Hydra computes calibration constants 19 and 20 and returns the calibrated reading. (For software versions lower than 5.4, this step computes calibration constants 17 and 18.)

4-30. Updating 2635A Data Bucket Embedded Instrument Firmware

The instrument firmware in the 2635A Hydra Data Bucket can be easily updated without even opening the instrument case or replacing any parts. The instrument firmware is stored in electrically eraseable and programmable Flash memory.

A diskette which contains the necessary software and the latest release of 2635A Data Bucket firmware may be obtained from either your local Fluke authorized service center or the Fluke factory. The local service centers are listed in Section 6 of this manual. Contact the one nearest you. To request the "2635A Embedded Firmware Memory Loader" diskette from the factory, telephone or send a fax to:

Fluke Corporation, Data Acquisition Sales Support Telephone: (206) 356-5870 FAX: (206) 356-5790

4-27

Page 167
Image 167
Fluke 2625A Updating 2635A Data Bucket Embedded Instrument Firmware, Wire Ohms Calibration Fixed Resistor Command

2635A, 2625A specifications

The Fluke 2625A is a sophisticated data logger designed for a wide range of industrial and laboratory applications. This versatile instrument excels in capturing and monitoring temperature and voltage measurements with precision and reliability. One of its standout features is its ability to record data from various sensors, making it ideal for environments where comprehensive data acquisition is critical.

Equipped with a large LCD display, the Fluke 2625A provides clear and user-friendly access to real-time measurement data. Its graphical interface allows users to visualize trends and monitor readings easily, which significantly enhances usability. The device supports a variety of input types and can connect to thermocouples, RTDs, and other sensors, offering great flexibility for various measurement tasks.

The Fluke 2625A employs advanced measurement technologies that ensure accuracy in data logging. With a high sampling rate, it captures fast-changing signals while maintaining data integrity. The device features a built-in memory that allows for extensive logging, accommodating long-term monitoring tasks without requiring constant oversight. Users can set it to log data at specified intervals, offering customizable solutions for different monitoring needs.

Another significant characteristic of the Fluke 2625A is its robust communication capabilities. The device supports USB and RS-232 interfaces for easy connectivity with computers and other devices, making data transfer and analysis seamless. This feature is especially beneficial for users who need to compile or analyze large sets of data quickly.

Furthermore, the Fluke 2625A is designed with durability in mind. Its rugged construction makes it suitable for challenging environments, including those with high levels of vibration, dust, or moisture. This reliability ensures that the device can withstand the rigors of fieldwork while still delivering accurate measurements.

In summary, the Fluke 2625A is a powerful data logger that combines versatility, accuracy, and reliability. Its user-friendly interface, advanced measurement technologies, and robust design make it an essential tool for professionals in various industries, from manufacturing and HVAC to research and development. Whether monitoring temperature changes in a laboratory or assessing voltage levels in an industrial setting, the Fluke 2625A stands out as a trusted solution for effective data logging and analysis.