3.Verify that Hydra is still in the total measuring mode. If not, press the TOTAL button. Reset the totalizer count shown on the display by pressing the SHIFT and TOTAL(ZERO) buttons.

The Hydra display should now show a value of 0.

4.Connect the output of the signal generator to the and J terminals.

5.Program the signal generator to output a 1.5V rms sine signal at 10 Hz.

The Hydra display should now show the totalizing value incrementing at a 10 count per second rate.

4-18. Dedicated Alarm Output Test

The Dedicated Alarm Output Test verifies that Alarm Outputs 0 through 3 are functioning properly. Because this test is dependent on voltage readings the Accuracy Verification Test for channel 0 and the Channel Integrity Test for channels 1 through 3 should be performed if voltage readings are suspect.

1.Switch OFF power to the instrument and disconnect all high voltage inputs.

2.Remove the eight-terminal Alarm Output connector from the rear of Hydra and all external connections to it. Connect short wires (to be used as test leads) to the [J] and 0 through 3 terminals. Leave the other ends of the wires unconnected at this time. Reinstall the connector.

3.Remove the Input Module from the rear of Hydra. Open the Input Module and jumper the H (high) terminal of channels 1, 2, and 3 together. Connect a test lead to the H of channel 1. Also jumper the L (low) terminals of channel 1, 2, and 3 together. Connect a second test lead to the L of channel 1. Reinstall the Input Module into Hydra. Refer to Figure 4-3.

4.Switch power ON.

5.Using a digital multimeter (DMM), verify that alarm outputs 0 through 3 are in the OFF or HIGH state. Perform this test by connecting the low or common of the multimeter to the ground test lead and the high of the multimeter to the alarm output. Verify a voltage greater than +3.8V dc.

6.Connect a cable from the Output VA HI and LO connectors of the 5700A to the VΩ and COM connectors on the front panel of Hydra. Then jumper the Hydra VΩ terminal to the H (high) test lead of the Input Module and the COM terminal to the L (low) test lead.

7.On Hydra, select the VDC function, 3V range, and assign a HI alarm limit of +1.0000 for channels 0 through 3. Set up all other channels (4-20) to the OFF function. Select a scan interval of 5 seconds.

8.Set the 5700A to output +0.9900 volts.

9.Press SCAN. Hydra should scan channels 0 through 3 every 5 seconds.

10.Using a digital multimeter, again verify that alarm outputs 0 through 3 are in the OFF or HIGH state.

11.Set the 5700A to output +1.1000 volts. Verify that the alarm outputs 0 through 3 are in the ON or LOW state (measure less than +0.8V dc).

4-16

Page 156
Image 156
Fluke 2635A, 2625A service manual Dedicated Alarm Output Test

2635A, 2625A specifications

The Fluke 2625A is a sophisticated data logger designed for a wide range of industrial and laboratory applications. This versatile instrument excels in capturing and monitoring temperature and voltage measurements with precision and reliability. One of its standout features is its ability to record data from various sensors, making it ideal for environments where comprehensive data acquisition is critical.

Equipped with a large LCD display, the Fluke 2625A provides clear and user-friendly access to real-time measurement data. Its graphical interface allows users to visualize trends and monitor readings easily, which significantly enhances usability. The device supports a variety of input types and can connect to thermocouples, RTDs, and other sensors, offering great flexibility for various measurement tasks.

The Fluke 2625A employs advanced measurement technologies that ensure accuracy in data logging. With a high sampling rate, it captures fast-changing signals while maintaining data integrity. The device features a built-in memory that allows for extensive logging, accommodating long-term monitoring tasks without requiring constant oversight. Users can set it to log data at specified intervals, offering customizable solutions for different monitoring needs.

Another significant characteristic of the Fluke 2625A is its robust communication capabilities. The device supports USB and RS-232 interfaces for easy connectivity with computers and other devices, making data transfer and analysis seamless. This feature is especially beneficial for users who need to compile or analyze large sets of data quickly.

Furthermore, the Fluke 2625A is designed with durability in mind. Its rugged construction makes it suitable for challenging environments, including those with high levels of vibration, dust, or moisture. This reliability ensures that the device can withstand the rigors of fieldwork while still delivering accurate measurements.

In summary, the Fluke 2625A is a powerful data logger that combines versatility, accuracy, and reliability. Its user-friendly interface, advanced measurement technologies, and robust design make it an essential tool for professionals in various industries, from manufacturing and HVAC to research and development. Whether monitoring temperature changes in a laboratory or assessing voltage levels in an industrial setting, the Fluke 2625A stands out as a trusted solution for effective data logging and analysis.