Figure 3-3. Removing the Handle and Handle Mounting Brackets

Note

The Display PCA provides a space for a center securing screw. If the two tabs are intact, this screw is not necessary. If a tab is broken, a screw can be used as an additional securing device.

The elastomeric Keypad Assembly (L) can now be lifted away from the Front Panel Assembly.

Only if necessary, gently remove the display window (M) by releasing the two snaps along its inside, bottom edge. While pushing slightly on the rear of the window, gently lever each snap by pressing against an adjacent edge on the keypad housing.

Caution

Avoid using ammonia or methyl-alcohol cleaningagents on either the Front Panel or the displaywindow. These types of cleaners can damage surfacefeatures and markings. Use an isopropyl-basedcleaning agent or water to clean the Front Paneland the display window.