HYDRA

Service Manual

Table 4-2. Performance Tests (Voltage, Resistance, and Frequency) (cont)

FUNCTION

RANGE

INPUT

 

 

LEVEL

 

 

 

300Ω

short

 

100Ω

 

3 kΩ

short

 

3 kΩ

1 kΩ

 

30 kΩ

10 kΩ

 

300 kΩ

100 kΩ

 

3 MΩ

1 MΩ

 

10 MΩ

10

* Optional test point if standards available.

FREQUENCY

DISPLAY ACCURACY

(1 Year, 18-28°C)

MINMAX

0.000.09

99.92100.15

0.00000.0003

0.99921.0009

9.99210.008

99.92100.08

0.99921.0008

9.98610.014

Note

All channels (0 through 20) can accommodate 2-terminal resistance measurements. Channel 0, with only two connections, cannot be used for 4- terminal measurements. Four-terminal resistance measurements can be defined for channels 1 through 10 only. Channels 11 through 20 are used, as required, for 4-terminal to provide the additional two connections. For example, a 4-terminal set up on channel 1 uses channels 1 and 11, each channel providing two connections.

Frequency

90 kHz

10 kHz/2V p-p

9.994

10.006

8.Connect a cable from the Output VA HI and LO of the 5700A to the Input Module test leads (observe proper polarity).

9.Select the VDC function and 300-volt range on Hydra and apply 0V dc from the 5700A. Then apply 290V dc input from the 5700A. Ensure the display reads between the minimum and maximum values as shown in Table 4-2 for the 0 and 290V dc input levels.

Note

Channels 0, 1, and 11 can accommodate a maximum input of 300V dc or ac. However, the maximum input for all other channels can only be 150V dc or ac.

10.With the exception of the selected voltage range and input voltagefrom the 5700A, repeat steps 1 through 9 for each remaining InputModule channel (2 through 20). Channels 2 through 10 and 12 through20 should be configured for the 150V dc range and an input voltageof 150 volts.

4-6. Thermocouple Measurement Range Accuracy Test

Verify that the Accuracy Verification Test for channel 0 meets minimum acceptable levels before performing this test.

Thermocouple temperature measurements are accomplished using the Hydra internal 100 mV and 1V dc ranges. (The ranges are not configurable by the operator.) This procedure provides the means to test these ranges.

Testing the 100 mV and 1V dc ranges requires computer interfacing with a host (terminal or computer). The host must send commands to select these ranges. These ranges cannot be selected from the Hydra front panel.

4-6

Page 146
Image 146
Fluke 2635A, 2625A service manual Thermocouple Measurement Range Accuracy Test, Function Range Input Level

2635A, 2625A specifications

The Fluke 2625A is a sophisticated data logger designed for a wide range of industrial and laboratory applications. This versatile instrument excels in capturing and monitoring temperature and voltage measurements with precision and reliability. One of its standout features is its ability to record data from various sensors, making it ideal for environments where comprehensive data acquisition is critical.

Equipped with a large LCD display, the Fluke 2625A provides clear and user-friendly access to real-time measurement data. Its graphical interface allows users to visualize trends and monitor readings easily, which significantly enhances usability. The device supports a variety of input types and can connect to thermocouples, RTDs, and other sensors, offering great flexibility for various measurement tasks.

The Fluke 2625A employs advanced measurement technologies that ensure accuracy in data logging. With a high sampling rate, it captures fast-changing signals while maintaining data integrity. The device features a built-in memory that allows for extensive logging, accommodating long-term monitoring tasks without requiring constant oversight. Users can set it to log data at specified intervals, offering customizable solutions for different monitoring needs.

Another significant characteristic of the Fluke 2625A is its robust communication capabilities. The device supports USB and RS-232 interfaces for easy connectivity with computers and other devices, making data transfer and analysis seamless. This feature is especially beneficial for users who need to compile or analyze large sets of data quickly.

Furthermore, the Fluke 2625A is designed with durability in mind. Its rugged construction makes it suitable for challenging environments, including those with high levels of vibration, dust, or moisture. This reliability ensures that the device can withstand the rigors of fieldwork while still delivering accurate measurements.

In summary, the Fluke 2625A is a powerful data logger that combines versatility, accuracy, and reliability. Its user-friendly interface, advanced measurement technologies, and robust design make it an essential tool for professionals in various industries, from manufacturing and HVAC to research and development. Whether monitoring temperature changes in a laboratory or assessing voltage levels in an industrial setting, the Fluke 2625A stands out as a trusted solution for effective data logging and analysis.