HYDRA

Service Manual

7.The RTD Temperature Accuracy test is complete. However, if you desire to perform this test on Input Module channels (2 through 10), repeat steps 1 through 5 substituting in the appropriate channel number.

Note

The only type of temperature measurement that can be made on channel 0 is 2-terminal RTD. Channels 11 through 20 support only 2-terminal RTDs.

Table 4-5. Performance Tests for RTD Temperature Function (Resistance Source)

Decade Reisitance

Source

100Ω

200Ω

300Ω

Simulated °C Temperature

2620A/2635A*

2635A*

 

 

0

0

266.58266.34

558.00557.70

Temperature Accuracy Specifications

1 Year @ 18-28°C

±0.24°C

±0.48°C

±0.75°C

*RTD temperature linearizations changed between the 2620A/2625A and 2635A Hydra instrument firmware. The 2620A & 2625A Hydra instruments are based on the International Practical Temperature Scale of 1968 (IPTS-68). The 2635A Hydra Instrument is based on the International Temperature Scale of 1990 (ITS-90).

These figures assume that RTD R0 is set to 100.00Ω for each channel. Accuracy given is for 4-wire measurements only.

4-12. RTD Temperature Accuracy Test (Using DIN/IEC 751).

1.Switch OFF power to the instrument and disconnect all other high voltage inputs.

2.Remove the Input Module from the rear of the instrument. Open the Input Module and connect a Platinum RTD, conforming to the European Standards IEC 751 (DIN 43760).

2-terminal RTD: Connect the RTD excitation leads to the H (high) and L (low) terminals of channel 1.

4-terminal RTD: Connect the RTD excitation leads (one red and one black wire) to the H (high) and L (low) terminals of channel 11. Connect the second pair of RTD red and black leads to the H and L leads of channel 1. (Refer to Figures 4-1 and 4-2 for proper connection.) Reinstall the Input Module into the instrument.

Note

4-terminal connections are made using pairs of channels. 4-terminal measurements can only be made on channels 1 through 10. Their accompanying pairs are channels 11 through 20.

3.Switch the instrument ON.

4.Insert the RTD probe and a mercury thermometer in a room temperature bath. Allow 20 minutes for thermal stabilization.

5.Depending on the type of connection made in step 2, select either the 2-Terminal or 4-Terminal RTD temperature function, RTD type PT (DIN IEC 751), for channel 1 on Hydra. Press MON and ensure the display reads the temperature of the room temperature bath (within tolerances shown in Table 4-6) as measured by the mercury thermometer.

4-12

Page 152
Image 152
Fluke 2625A service manual RTD Temperature Accuracy Test Using DIN/IEC, Simulated C Temperature 2620A/2635A

2635A, 2625A specifications

The Fluke 2625A is a sophisticated data logger designed for a wide range of industrial and laboratory applications. This versatile instrument excels in capturing and monitoring temperature and voltage measurements with precision and reliability. One of its standout features is its ability to record data from various sensors, making it ideal for environments where comprehensive data acquisition is critical.

Equipped with a large LCD display, the Fluke 2625A provides clear and user-friendly access to real-time measurement data. Its graphical interface allows users to visualize trends and monitor readings easily, which significantly enhances usability. The device supports a variety of input types and can connect to thermocouples, RTDs, and other sensors, offering great flexibility for various measurement tasks.

The Fluke 2625A employs advanced measurement technologies that ensure accuracy in data logging. With a high sampling rate, it captures fast-changing signals while maintaining data integrity. The device features a built-in memory that allows for extensive logging, accommodating long-term monitoring tasks without requiring constant oversight. Users can set it to log data at specified intervals, offering customizable solutions for different monitoring needs.

Another significant characteristic of the Fluke 2625A is its robust communication capabilities. The device supports USB and RS-232 interfaces for easy connectivity with computers and other devices, making data transfer and analysis seamless. This feature is especially beneficial for users who need to compile or analyze large sets of data quickly.

Furthermore, the Fluke 2625A is designed with durability in mind. Its rugged construction makes it suitable for challenging environments, including those with high levels of vibration, dust, or moisture. This reliability ensures that the device can withstand the rigors of fieldwork while still delivering accurate measurements.

In summary, the Fluke 2625A is a powerful data logger that combines versatility, accuracy, and reliability. Its user-friendly interface, advanced measurement technologies, and robust design make it an essential tool for professionals in various industries, from manufacturing and HVAC to research and development. Whether monitoring temperature changes in a laboratory or assessing voltage levels in an industrial setting, the Fluke 2625A stands out as a trusted solution for effective data logging and analysis.