Diagnostic Testing and Troubleshooting (2620A/2625A)

5

Digital and Alarm Output Troubleshooting

5-15. Digital and Alarm Output Troubleshooting

Power up Hydra while holding down the CANCL button to reset the instrument configuration. Since the structure of the eight Digital Outputs and four Alarm Outputs is very similar, the troubleshooting procedure presented here does not refer to specific device and pin numbers. First verify that the input of the Output Driver (A1U17 or A1U27) is low and that the output is near +5V dc. If the input is high, the problem may be in the address decoding (A1U12 and A1U15) or the associated octal D-type flip-flop (A1U16 or A1U26). If the output is not near +5V dc, use an ohmmeter to check the pull- up resistor in A1Z2.

Use the proper computer interface command to change the state of the Digital Output (DO_LEVEL x,1) or Alarm Output (ALARM_DO_LEVEL x,1), where x is the number of the output being checked. Now verify that the input of the Output Driver is high and that the output is near +0.8V dc. If there is no change in the input, check the address decoding and operation of the associated octal D-type flip-flop (A1U16 or A1U26.) If the output failed to change, the problem is most likely the inverting output driver (A1U17 or A1U27).

5-16. Digital Input Troubleshooting

Power up Hydra while holding down the CANCL button to reset the instrument configuration. Verify that the Input Buffer Threshold circuit generates approximately 1.4V dc at A1TP18. Drive the Digital Input (A1J5) to be checked with a signal generator outputting a 100-Hz square wave that transitions from 0 to +5V dc. The signal generator output common should be connected to Common (A1J5-1). Verify that the output of the Input Buffer is a 100 Hz square wave that is the inverse of the input signal.

If the Input Buffer does not function correctly, the problem is probably A1Z1, A1Z3, or the associated comparator (A1AR2 or A1AR3). If the Input Buffer functions correctly, but Hydra is not able to read the state of the Digital Input correctly, the problem is most likely the tri-state buffer A1U13. If Hydra is not able to read the states of any of the eight Digital Inputs correctly, the problem is most likely the address decoding (A1U10 and A1U12) for the tri-state buffer.

5-17. Totalizer Troubleshooting

Power up Hydra while holding down the CANCL button to reset the instrument configuration. Verify that the Input Buffer Threshold circuit generates approximately 1.4V dc at A1TP18. Drive the Totalizer Input (A1J5-2) with a signal generator outputting a 100-Hz square wave that transitions from 0 to +5V dc. The signal generator output common should be connected to Common (A1J5-1). Verify that the output of the Input Buffer (A1AR1-7) is a 100-Hz square wave that is the inverse of the input signal. Verify also that the input to the totalizer counter (A1TP20) is a buffered form of the signal just verified at the output of the Input Buffer.

Use the following procedure to troubleshoot the totalizer input debouncer, Enable the totalizer debouncer by sending the TOTAL_DBNC 1 Computer Interface command to the instrument; verify that A1U16-16 is now high. With the signal generator still connected and outputting a 100-Hz square wave, verify that A1U14 drives the clear input of the counter (A1U20-11) low for 1.67 milliseconds after each edge of the input signal. Verify that counter output A1U20-13 generates a 4.8-kHz clock while A1U20-11 is low. Verify that the shift register output (A1U29-9) changes to the same state as the totalizer input signal about 1.67 milliseconds after a transition occurs on the totalizer input signal (A1U29-10).

5-21

Page 191
Image 191
Fluke 2625A, 2635A Digital and Alarm Output Troubleshooting, Digital Input Troubleshooting, Totalizer Troubleshooting

2635A, 2625A specifications

The Fluke 2625A is a sophisticated data logger designed for a wide range of industrial and laboratory applications. This versatile instrument excels in capturing and monitoring temperature and voltage measurements with precision and reliability. One of its standout features is its ability to record data from various sensors, making it ideal for environments where comprehensive data acquisition is critical.

Equipped with a large LCD display, the Fluke 2625A provides clear and user-friendly access to real-time measurement data. Its graphical interface allows users to visualize trends and monitor readings easily, which significantly enhances usability. The device supports a variety of input types and can connect to thermocouples, RTDs, and other sensors, offering great flexibility for various measurement tasks.

The Fluke 2625A employs advanced measurement technologies that ensure accuracy in data logging. With a high sampling rate, it captures fast-changing signals while maintaining data integrity. The device features a built-in memory that allows for extensive logging, accommodating long-term monitoring tasks without requiring constant oversight. Users can set it to log data at specified intervals, offering customizable solutions for different monitoring needs.

Another significant characteristic of the Fluke 2625A is its robust communication capabilities. The device supports USB and RS-232 interfaces for easy connectivity with computers and other devices, making data transfer and analysis seamless. This feature is especially beneficial for users who need to compile or analyze large sets of data quickly.

Furthermore, the Fluke 2625A is designed with durability in mind. Its rugged construction makes it suitable for challenging environments, including those with high levels of vibration, dust, or moisture. This reliability ensures that the device can withstand the rigors of fieldwork while still delivering accurate measurements.

In summary, the Fluke 2625A is a powerful data logger that combines versatility, accuracy, and reliability. Its user-friendly interface, advanced measurement technologies, and robust design make it an essential tool for professionals in various industries, from manufacturing and HVAC to research and development. Whether monitoring temperature changes in a laboratory or assessing voltage levels in an industrial setting, the Fluke 2625A stands out as a trusted solution for effective data logging and analysis.