Diagnostic Testing and Troubleshooting (2620A/2625A)

5

Digital Kernel Troubleshooting

5-14. Digital Kernel Troubleshooting

At power-up, if the display does not light or lights up and fails to report errors or begin operation, use the following troubleshooting procedures.

First check the state of SWR1 (A1U4-21). If this status line is less than 0.8V, basic processor operation is intact. Examining SWR2 through SWR5 (A1U4-22 through -25, respectively) should indicate how far the software progressed before finding an error. If the state of SWR1 is not less than 0.8V, the problem may be in the 6303Y Microprocessor (A1U4), the ROM or NVRAM decode circuitry (A1U10 and A1U21), the ROM (A1U8) or NVRAM (A1U3), or the address/data lines among these parts.

Note

The functions of SWR1 through SWR5 as power-upstatus lines persist for only 3 to 4 seconds. Thesefunctions end when the keyboard scanner beginsoperation (if it can). Extremely difficult casesmay require the use of an oscilloscope triggered onthe falling edge of SWR1 to examine the states ofSWR2 through SWR5.

To determine the relative health of the 6303Y Microprocessor (A1U4), first check for a valid E clock at pin 68. The default for the E clock after reset is a rectangular wave with a period of 1.221 μs and a duty cycle of about 67%.

If the processor is able to fetch instructions from the ROM, the software initializes the processor, and the E clock becomes a square wave with a period of 0.814 us. Since this initialization occurs almost immediately with a working instrument, the resulting square wave on the E clock line is a good indication that the software has begun to execute.

If the E clock remains a 1.221 μs rectangular wave, the SWR2 (A1U4-22) keyboard scan line might be shorted to ground. This condition would cause the Microprocessor to HALT after reset. Check whether the 6303Y Microprocessor is attempting to access ROM; LIR* (A1U4-64) should transition for a short period of time after reset. If it does, the 6303Y Microprocessor is probably operational, and the problem is external to the processor.

The processor can execute an instruction that stops both itself and the E clock. Therefore, the absence of any activity on pin 68 does not necessarily mean that A1U4 or A1Y1 is bad. If some other failure prevents proper ROM access, the processor may have just "gone to sleep". This can be verified by checking for a rectangular wave occurring at pin 68 for a short time after RESET* transitions high on pin 7. A1U4 and A1Y1 are probably operational if this rectangular wave is at least momentarily present.

To check the ROM decode circuitry, verify that A1U10-6 is transitioning low and that these transitions correspond roughly to the low-going transitions of LIR*. Pin 6 must be low when LIR* is low. Verify that this signal also appears at the ROM Chip Enable, A1U8-20. If the ROM Chip Enable is present, the problem is with the ROM itself or there is a fault in the address/data lines among the 6303Y Microprocessor, ROM, NVRAM, and Option Connector.

If SWR1 (A1U4-21) and SWR2 (A1U4-22) transition low, but SWR3 (A1U4-23) remains high, the problem is with the NVRAM decode circuitry (A1U15, A1U21), the external NVRAM (A1U3), or the address/data/control lines between the NVRAM and the 6303Y Microprocessor.

To check the NVRAM decode circuitry, verify that A1U21-6 is transitioning low and that these transitions correspond approximately to the low-going transitions of WR* (A1U4-66). It may be necessary to continually reset (power on) the instrument to check these lines, since the activity probably halts quickly when the instrument software goes awry. Verify that the signal on A1U21-6 also appears at the NVRAM Chip Enable, A1U3-20. If the NVRAM Chip Enable is present, the problem is with either the NVRAM itself or the address, data, RD*, or WR* lines between the 6303Y Microprocessor and the external NVRAM.

5-19

Page 189
Image 189
Fluke 2625A, 2635A service manual Digital Kernel Troubleshooting

2635A, 2625A specifications

The Fluke 2625A is a sophisticated data logger designed for a wide range of industrial and laboratory applications. This versatile instrument excels in capturing and monitoring temperature and voltage measurements with precision and reliability. One of its standout features is its ability to record data from various sensors, making it ideal for environments where comprehensive data acquisition is critical.

Equipped with a large LCD display, the Fluke 2625A provides clear and user-friendly access to real-time measurement data. Its graphical interface allows users to visualize trends and monitor readings easily, which significantly enhances usability. The device supports a variety of input types and can connect to thermocouples, RTDs, and other sensors, offering great flexibility for various measurement tasks.

The Fluke 2625A employs advanced measurement technologies that ensure accuracy in data logging. With a high sampling rate, it captures fast-changing signals while maintaining data integrity. The device features a built-in memory that allows for extensive logging, accommodating long-term monitoring tasks without requiring constant oversight. Users can set it to log data at specified intervals, offering customizable solutions for different monitoring needs.

Another significant characteristic of the Fluke 2625A is its robust communication capabilities. The device supports USB and RS-232 interfaces for easy connectivity with computers and other devices, making data transfer and analysis seamless. This feature is especially beneficial for users who need to compile or analyze large sets of data quickly.

Furthermore, the Fluke 2625A is designed with durability in mind. Its rugged construction makes it suitable for challenging environments, including those with high levels of vibration, dust, or moisture. This reliability ensures that the device can withstand the rigors of fieldwork while still delivering accurate measurements.

In summary, the Fluke 2625A is a powerful data logger that combines versatility, accuracy, and reliability. Its user-friendly interface, advanced measurement technologies, and robust design make it an essential tool for professionals in various industries, from manufacturing and HVAC to research and development. Whether monitoring temperature changes in a laboratory or assessing voltage levels in an industrial setting, the Fluke 2625A stands out as a trusted solution for effective data logging and analysis.