Theory of Operation (2635A) 2A

Detailed Circuit Description

+ REFERENCE

 

 

 

(– INPUT)

 

 

 

+

 

 

+

 

 

 

REFERENCE

 

 

COUNTER

 

 

 

_

_

 

A3C13

A/D

 

 

 

COMPARATOR

 

INTEGRATE

S77

 

 

 

 

–REFERENCE

REFERENCE

 

 

 

 

 

(+ INPUT)

 

 

 

 

+

A3Z2

 

 

 

 

INPUT HI

_

_

 

 

 

 

BUFFER

+

 

 

 

INTEGRATOR

 

INPUT LO

INTEGRATE

 

 

INPUT

 

 

 

 

 

 

 

 

s17f.eps

Figure 2A-7. A/D Converter Simplified Schematic (2635A)

In both the slow and fast measurement rates, the a/d converter uses its ±300 mV range for most measurement functions and ranges. The primary exceptions are that the 3V dc range is measured on the a/d converter 3V range, thermocouples are measured on the ±100 mV range, and the temperature reference is measured on the 1V a/d converter range. The typical overload point on a slow rate 30000 count range is 32000 display counts; the typical overload point on a fast rate 3000 count range is 3200 display counts. During the integrate phase, the a/d buffer in the A3U8 Analog Measurement Processor applies the signal to be measured to one of the four integrator input resistors in network A3Z2. As shown on the A/D Converter schematic diagram in Section 8, the choice of resistor selects the a/d converter range. Switch S69 connects the buffer output through pin B.1 for the 100-mV range, S71 connects the output through B.32 for the 300 mV range, S73 connects to pin B1 for the 1V range, and S75 sets up the 3V range through pin B3.2.

The current through the selected integrator input resistor charges integrator capacitor A3C13, with the current dependent on the buffer output voltage. After the integrate phase, the buffer is connected to the opposite polarity reference voltage, and the integrator integrates back toward zero capacitor voltage until the comparator trips. An internal counter measures this variable integrate time. If the a/d converter input voltage is too high, the integrator overloads and does not return to its starting point by the end of the measurement phase. Switch S77 is then turned on to discharge integrate capacitor A3C13.

The reference voltage used during the variable integrate period for voltage (and high ohms) conversions is generated from zener reference diode A3VR1, which is time and temperature stable. The reference amplifier in the Analog Measurement Processor, along with resistors A3R15, A3R18, and A3R21, pulls approximately 2 mA of current through the zener. Resistors in network A3Z2 divide the zener voltage down to the reference 1.05V required by the A/D Converter.

2A-31

Page 113
Image 113
Fluke 2625A, 2635A service manual 2A-31

2635A, 2625A specifications

The Fluke 2625A is a sophisticated data logger designed for a wide range of industrial and laboratory applications. This versatile instrument excels in capturing and monitoring temperature and voltage measurements with precision and reliability. One of its standout features is its ability to record data from various sensors, making it ideal for environments where comprehensive data acquisition is critical.

Equipped with a large LCD display, the Fluke 2625A provides clear and user-friendly access to real-time measurement data. Its graphical interface allows users to visualize trends and monitor readings easily, which significantly enhances usability. The device supports a variety of input types and can connect to thermocouples, RTDs, and other sensors, offering great flexibility for various measurement tasks.

The Fluke 2625A employs advanced measurement technologies that ensure accuracy in data logging. With a high sampling rate, it captures fast-changing signals while maintaining data integrity. The device features a built-in memory that allows for extensive logging, accommodating long-term monitoring tasks without requiring constant oversight. Users can set it to log data at specified intervals, offering customizable solutions for different monitoring needs.

Another significant characteristic of the Fluke 2625A is its robust communication capabilities. The device supports USB and RS-232 interfaces for easy connectivity with computers and other devices, making data transfer and analysis seamless. This feature is especially beneficial for users who need to compile or analyze large sets of data quickly.

Furthermore, the Fluke 2625A is designed with durability in mind. Its rugged construction makes it suitable for challenging environments, including those with high levels of vibration, dust, or moisture. This reliability ensures that the device can withstand the rigors of fieldwork while still delivering accurate measurements.

In summary, the Fluke 2625A is a powerful data logger that combines versatility, accuracy, and reliability. Its user-friendly interface, advanced measurement technologies, and robust design make it an essential tool for professionals in various industries, from manufacturing and HVAC to research and development. Whether monitoring temperature changes in a laboratory or assessing voltage levels in an industrial setting, the Fluke 2625A stands out as a trusted solution for effective data logging and analysis.