Diagnostic Testing and Troubleshooting (2635A) 5A

Digital and Alarm Output Troubleshooting

Figure 5A-7 shows the timing relationships of the MC68302 Microprocessor address, data, and memory control signals used for memory read and write cycles. The chip selects from the Microprocessor (FLASH*, SRAM*, XMCARD*, and I/O*) are decoded internally from the address bus and the address strobe (AS*) signal. Therefore the AS* waveform is the same as the chip select signal for the device that the Microprocessor is accessing.

If the instrument powers up without any errors, but does not recognize front-panel button presses, the problem may be in the Keyboard Interrupt (KINT*) signal from the FPGA. If the KINT* output (A1U25-62) is low (and never goes high), the Microprocessor (A1U1) is failing to recognize the interrupt, or the microprocessor interface to A1U25 is not working correctly. Verify that the KINT* signal gets to the input pin (A1U1-121) on the Microprocessor.

If the interval time does not count down properly when scanning is enabled, the problem may be in the Real-Time Clock Interrupt (CINT*) signal from the Real-Time Clock (A1U12). If the CINT* output (A1U12-3) is low (and never goes high), the Microprocessor (A1U1) is failing to recognize the interrupt or the microprocessor interface to A1U12 is not working correctly. Verify that the CINT* signal gets to the input pin (A1U1-96) on the Microprocessor. Also verify that the Real-Time Clock is actively keeping time by checking for a 1-Hz square wave output on A1U12-4. Pin A1U12-2 must also be properly connected to GND for the Real-Time Clock to operate. The interface to the Microprocessor operates very much like the interface to NVRAM device A1U20.

5A-15. Digital and Alarm Output Troubleshooting

Power up Hydra while holding down the CANCL button to reset the instrument configuration. Since the structure of the eight Digital Outputs and four Alarm Outputs is very similar, the troubleshooting procedure presented here does not refer to specific device and pin numbers. First verify that the input of the Output Driver (A1U17 or A1U27) is low and that the output is near +5V dc. If the input is high, the problem may be in the FPGA (A1U25). If the output is not near +5V dc, use an ohmmeter to check the pull-up resistor in A1Z2.

Use the proper computer interface command to change the state of the Digital Output (DO_LEVEL x,1) or Alarm Output (ALARM_DO_LEVEL x,1), where x is the number of the output being checked. Now verify that the input of the Output Driver is high and that the output is near +0.8V dc. If there is no change in the input, check the address signals to the FPGA (A1U25-85, A1U25-90, A1U25-96, A1U25-97) and the behavior of the output pin on the FPGA that goes to the input of the Output Driver (A1U17 or A1U27). If the output of the Output Driver fails to change when the input does, the problem is most likely the inverting output driver (A1U17 or A1U27).

5A-16. Digital Input Troubleshooting

Power up Hydra while holding down the CANCL button to reset the instrument configuration. Verify that the Input Buffer Threshold circuit generates approximately 1.4V dc at A1TP18. Drive the Digital Input (A1J5) to be checked with a signal generator outputting a 100-Hz square wave that transitions from 0 to +5V dc. The signal generator output common should be connected to Common (A1J5-1). Verify that the output of the Input Buffer is a 100-Hz square wave that is the inverse of the input signal.

5A-21

Page 221
Image 221
Fluke 2625A, 2635A service manual 5A-15. Digital and Alarm Output Troubleshooting, 5A-16. Digital Input Troubleshooting

2635A, 2625A specifications

The Fluke 2625A is a sophisticated data logger designed for a wide range of industrial and laboratory applications. This versatile instrument excels in capturing and monitoring temperature and voltage measurements with precision and reliability. One of its standout features is its ability to record data from various sensors, making it ideal for environments where comprehensive data acquisition is critical.

Equipped with a large LCD display, the Fluke 2625A provides clear and user-friendly access to real-time measurement data. Its graphical interface allows users to visualize trends and monitor readings easily, which significantly enhances usability. The device supports a variety of input types and can connect to thermocouples, RTDs, and other sensors, offering great flexibility for various measurement tasks.

The Fluke 2625A employs advanced measurement technologies that ensure accuracy in data logging. With a high sampling rate, it captures fast-changing signals while maintaining data integrity. The device features a built-in memory that allows for extensive logging, accommodating long-term monitoring tasks without requiring constant oversight. Users can set it to log data at specified intervals, offering customizable solutions for different monitoring needs.

Another significant characteristic of the Fluke 2625A is its robust communication capabilities. The device supports USB and RS-232 interfaces for easy connectivity with computers and other devices, making data transfer and analysis seamless. This feature is especially beneficial for users who need to compile or analyze large sets of data quickly.

Furthermore, the Fluke 2625A is designed with durability in mind. Its rugged construction makes it suitable for challenging environments, including those with high levels of vibration, dust, or moisture. This reliability ensures that the device can withstand the rigors of fieldwork while still delivering accurate measurements.

In summary, the Fluke 2625A is a powerful data logger that combines versatility, accuracy, and reliability. Its user-friendly interface, advanced measurement technologies, and robust design make it an essential tool for professionals in various industries, from manufacturing and HVAC to research and development. Whether monitoring temperature changes in a laboratory or assessing voltage levels in an industrial setting, the Fluke 2625A stands out as a trusted solution for effective data logging and analysis.