FUNCtion

[SENSe:]FUNCtion “< function>” selects the measurement function. You can select the functions shown in the following table.

Parameters

 

 

 

 

Parameter Name

Parameter Type

Range of Values

Default Units

 

 

 

 

<function>

discrete

:CURRent:AC

none

 

 

:CURRent[:DC]

 

 

 

:FREQuency

 

 

 

:FRESistance

 

 

 

:PERiod

 

 

 

:RESistance

 

 

 

:VOLTage:AC

 

 

 

:VOLTage[:DC]

 

 

 

:VOLTage[:DC]:RATio

 

 

 

 

 

 

 

 

 

Comments

*RST Condition: SENS:VOLT:DC

 

Example

Change Measurement Function

 

 

CONF:VOLT

Function: DC voltage.

 

FUNC “FRES”

Set function to 4-wire resistance.

 

READ?

Place multimeter in wait-for-trigger state and

 

 

make measurement; send reading to output

 

enter statement

buffer.

 

Enter reading into computer.

FUNCtion?

[SENSe:]FUNCtion? returns one of the following quoted strings to the output buffer:

CURR:AC” “ CURR

FREQ

FRES

PER

RES

VOLT:AC” “ VOLT

VOLT:RAT

Example Query the Measurement Function

FUNC “FRES”

Function: 4-wire ohms.

FUNC?

Query multimeter to return selected function.

enter statement

Enter quoted string into computer.

118 Multimeter Command Reference

Chapter 3

Page 118
Image 118