FREQuency:VOLTage:RANGe

[SENSe:]FREQuency:VOLTage:RANGe <range> selects the voltage range for the

signal level of frequency measurements.

Parameters

 

 

 

 

Parameter Name

Parameter Type

Range of Values

Default Units

 

 

 

 

<range>

numeric

100mV1V10V100V300V

volts

 

 

MINMAX

 

 

 

 

 

 

 

 

 

Comments To select a standard measurement range, specify range as the input signal’s maximum expected voltage. The multimeter then selects the correct range.

MIN selects the minimum range available with the VOLTage :RANGe command: 100mV. MAX selects the maximum range available: 300V.

Specifying a fixed range disables the autorange mode set by the

FREQ:VOLT:RANG:AUTO command.

The FREQ:VOLT:RANG command overrides the range setting from a previous

CONFigure:FREQuency command.

*RST Condition: FREQ:VOLT:RANG 10

Example Set the Voltage Range for Frequency Measurements to 100V

FREQ:VOLT:RANG 100

Voltage range for frequency measurements is

 

100V.

FREQuency:VOLTage:RANGe?

[SENSe:]FREQuency:VOLTage:RANGe? [MINMAX] returns one of the following numbers to the output buffer: 0.1, 1, 10, 100 or 300.

The present voltage range selected if MIN or MAX is not specified.The minimum voltage range available (100mV) if MIN is specified.The maximum voltage range available (300V) if MAX is specified.Example Query the Measurement Range

FREQ:VOLT:RANG 10

Select 10 V range.

FREQ:VOLT:RANG?

Query the present range.

enter statement

Enter value into computer.

Chapter 3

Multimeter Command Reference 129