AC Characteristics

Accuracy Specifications ± (% of reading + % of range) [1]

 

 

 

 

 

 

 

 

 

 

 

 

Temperature

 

 

 

 

 

 

 

 

 

 

 

 

 

Coefficient

 

 

 

 

 

 

 

 

 

 

24 Hour [2]

90 Day

1 Year

0°C - 18°C

 

 

Function

Range [3]

 

 

Frequency

23°C ± 1°C

23°C ± 5°C

23°C ± 5°C

28°C - 55°C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

100.0000 mV

3 Hz-5 Hz

1.00 + 0.03

1.00 + 0.04

1.00 + 0.04

0.100 + 0.004

 

 

 

 

 

 

5

Hz-10 Hz

0.35 + 0.03

0.35 + 0.04

0.35 + 0.04

0.035 + 0.004

 

 

 

 

 

 

10 Hz-20 kHz

0.04 + 0.03

0.05 + 0.04

0.06 + 0.04

0.005 + 0.004

 

 

 

 

 

 

20 kHz-50 kHz

0.10 + 0.05

0.11 + 0.05

0.12 + 0.05

0.011 + 0.005

 

 

 

 

 

 

50 kHz-100 kHz

0.55 + 0.08

0.60 + 0.08

0.60 + 0.08

0.060 + 0.008

 

 

True RMS AC

 

 

 

100 kHz-300 kHz

5.00 + 0.50

5.00 + 0.50

5.00 + 0.50

0.200 + 0.020

 

 

Voltage [4]

 

 

 

 

 

 

 

 

 

 

 

 

 

1.000000 V

 

3

Hz-5 Hz

1.00 + 0.02

1.00 + 0.03

1.00 + 0.03

0.100 + 0.003

 

 

 

 

 

 

 

 

to

 

5

Hz-10 Hz

0.35 + 0.02

0.35 + 0.03

0.35 + 0.03

0.035 + 0.003

 

 

 

100.000V

 

10 Hz-20 kHz

0.04 + 0.02

0.05 + 0.03

0.06 + 0.03

0.005 + 0.003

 

 

 

[12]

 

20 kHz-50 kHz

0.10 + 0.04

0.11 + 0.05

0.12 + 0.05

0.011 + 0.005

 

 

 

 

 

 

50 kHz-100 kHz

0.55 + 0.08

0.60 + 0.08

0.60 + 0.08

0.060 + 0.008

 

 

 

 

 

 

100 kHz-300 kHz

5.00 + 0.50

5.00 + 0.50

5.00 + 0.50

0.200 + 0.020

 

 

 

 

 

 

 

 

 

 

 

 

 

 

300.000V

 

3

Hz-5 Hz

1.00 + 0.06

1.00 + 0.09

1.00 + 0.09

0.100 + 0.009

 

 

 

[12]

 

5

Hz-10 Hz

0.35 + 0.06

0.35 + 0.09

0.35 + 0.09

0.035 + 0.009

 

 

 

 

 

 

10 Hz-20 kHz

0.04 + 0.06

0.05 + 0.09

0.06 + 0.09

0.005 + 0.009

 

 

 

 

 

 

20 kHz-50 kHz

0.10 + 0.12

0.11 + 0.15

0.12 + 0.15

0.011 + 0.015

 

 

 

 

 

 

50 kHz-100 kHz [5]

0.55 + 0.24

0.60 + 0.24

0.60 + 0.24

0.060 + 0.024

 

 

 

 

 

 

100 kHz-300 kHz [5]

5.00 + 1.50

5.00 + 1.50

5.00 + 1.50

0.200 + 0.060

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1.000000 A

 

3

Hz-5 Hz

1.05 + 0.04

1.05 + 0.04

1.05 + 0.04

0.100 + 0.006

 

 

 

 

 

 

5

Hz-10 Hz

0.35 + 0.04

0.35 + 0.04

0.35 + 0.04

0.035 + 0.006

 

 

True RMS AC

 

 

 

10 Hz-1 kHz

0.15 + 0.04

0.15 + 0.04

0.15 + 0.04

0.015 + 0.006

 

 

Current [4]

 

 

 

1 kHz-5 kHz

0.40 + 0.04

0.40 + 0.04

0.40 + 0.04

0.015 + 0.006

 

 

 

 

 

 

 

 

 

 

 

 

 

 

3.00000 A

 

3

Hz-5 Hz

1.70 + 0.06

1.70 + 0.06

1.70 + 0.06

0.100 + 0.006

 

 

 

 

 

 

5

Hz-10 Hz

0.95 + 0.06

0.95 + 0.06

0.95 + 0.06

0.035 + 0.006

 

 

 

 

 

 

10 Hz-1 kHz

0.75 + 0.06

0.75 + 0.06

0.75 + 0.06

0.015 + 0.006

 

 

 

 

 

 

1 kHz-5kHz

1.00 + 0.06

1.00 + 0.06

1.00 + 0.06

0.015 + 0.006

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Additional Low Frequency Errors

 

 

 

Additional Crest Factor Errors (non-sinewave)

 

 

(% of reading)

 

 

 

 

 

 

 

 

[6]

 

 

 

 

Frequency

 

 

AC Filter

 

 

 

Crest Factor

Error (% or reading)

 

 

 

 

3 Hz

20 Hz

200 Hz

 

1-2

0.05%

 

 

 

10Hz-20Hz

0

0.74

--

 

 

2-3

0.15%

 

 

 

20Hz-40Hz

0

0.22

--

 

 

3-4

0.30%

 

 

 

40Hz-100Hz

0

0.06

0.73

 

 

4-5

0.40%

 

 

 

100Hz-200Hz

0

0.01

0.22

 

 

 

 

 

 

 

200Hz-1kHz

0

 

 

0

0.18

 

 

NOTE: Crest Factor is not specified for

 

 

>1kHz

0

 

 

0

0

 

 

non sinewave inputs <100Hz using the

 

 

 

 

 

 

 

 

 

 

 

slow (3Hz) AC filter. See note [6].

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

174 HP E1312A and HP E1412A Multimeter Specifications