SCPI Command Quick Reference

 

 

 

 

The following tables summarize SCPI commands for the HP E1312A B-size and

 

HP E1412A C-size 6½-Digit Multimeters.

 

 

 

 

Command

Description

 

 

 

 

 

 

ABORt

 

Place multimeter in idle state.

 

 

 

CALCulate

:AVERage:AVERage?

Query average value of the average.

 

:AVERage:COUNt?

Query average count

 

:AVERage:MAXimum?

Query average maximum.

 

:AVERage:MINimum?

Query average minimum.

 

:DB:REFerence <value> MINMAX

Set dB reference value

 

:DB:REFerence? [MINMAX]

Query dB reference value.

 

:DBM:REFerence <value> MINMAX

Set dBm reference value.

 

:DBM:REFerence? [MINMAX]

Query dBm reference value.

 

:FUNCtion AVERageDBDBM LIMitNULL

Set math function to calculate.

 

:FUNCtion?

Query math function set.

 

:LIMit:LOWer <value> MINMAX

Set lower limit value.

 

:LIMit:LOWer? [MINMAX]

Query lower limit value.

 

:LIMit:UPPer <value> MINMAX

Set upper limit value.

 

:LIMit:UPPer? [MINMAX]

Query upper limit value.

 

:NULL:OFFSet <value> MINMAX

Set null offset value.

 

:NULL:OFFSet? [MINMAX]

Query null offset value.

 

:STATe OFFON

Enable/disable math function state.

 

:STATe?

Query math function state.

 

 

 

CALibration

:COUNt?

Query number of cal operations.

 

:LFRequency 5060400

Sets line reference frequency.

 

:LFRequency? [MINMAX]

Query line reference frequency.

 

:SECure:CODE <new code>

Enters a new security code.

 

:SECure:STATe OFFON,<code>

Enables/disables the security code.

 

:SECure:STATe?

Queries the security state.

 

:STRing <quoted string>

Lets you store info about your calibration.

 

:STRing?

Queries the cal string.

 

:VALue <cal_value>

Sets the calibration value.

 

:VALue?

Queries the calibration value.

 

:ZERO:AUTO ONOFF

Enable/disable autozero mode.

 

:ZERO:AUTO?

Query autozero mode.

 

 

 

CALibration?

 

Initiates the calibration process using the cal

 

 

value set by CAL:VALue. The command returns a

 

 

value to indicate the calibration was successful.

 

 

 

 

 

 

Chapter 3

Multimeter Command Reference 167