PERiod:VOLTage:RANGe:AUTO

[SENSe:]PERiod:VOLTage:RANGe:AUTO <mode> enables or disables the autorange function for the signal level of period measurements.

Parameters

 

 

 

 

 

 

Parameter Name

Parameter Type

Range of Values

Default Units

 

 

 

 

 

 

<mode>

boolean

OFF0ON1

none

 

 

 

 

 

 

 

 

 

 

Comments

You can substitute decimal values for the OFF (“ 0”) and ON (“ 1”) parameters.

When autoranging is ON, the multimeter samples the input before each measurement and selects the appropriate range.

If you explicitly select a range using PERiod:VOLT:RANGe, autoranging is turned OFF.

Related Commands: CONFigure, FREQuency:VOLT:RANGe

*RST Condition: PER:VOLT:RANG:AUTO ON

Example Disable Autoranging

PER:VOLT:RANG:AUTO OFF

Disable autorange.

PERiod:VOLTage:RANGe:AUTO?

[SENSe:]PERiod:VOLTage:RANGe:AUTO? returns a number to show whether the autorange mode is enabled or disabled: “ 1” = ON, “ 0” = OFF. The number is sent to the output buffer.

Example Query the Autorange Mode

PER:VOLT:RANG:AUTO OFF

Disable autorange.

PER:VOLT:RANG:AUTO?

Query multimeter to return autorange mode.

enter statement

Enter value into computer.