RESistance:RESolution

[SENSe:]RESistance:RESolution <resolution> selects the resolution for 2-wire resistance measurements.

Parameters

 

 

 

 

Parameter Name

Parameter Type

Range of Values

Default Units

 

 

 

 

<resolution>

numeric

resolutionMINMAX

ohms

 

 

 

 

 

 

 

 

Comments MINimum selects the best resolution (the smallest value) for the selected range. MAXimum selects the worst resolution (the largest value) for the selected range.

You must select a range using RESistance:RANGe before specifying resolution. Also, only specify a resolution when making measurements on a fixed range. Otherwise, the resolution will change to correspond with the range selected during autoranging.

If autoranging is required, set the resolution using the MIN or MAX parameters.

If necessary to achieve the specified resolution, the multimeter will increase the integration time as needed. This command overrides the results of previously executed RESistance:NPLC or RESistance:APERture command (the last command executed has priority).

The RES:RESolution command overrides the resolution setting from a previous CONFigure:RESistance command.

Related Commands: CONFigure, RESistance:NPLC

*RST Condition: Based on the *RST values for the RESistance:NPLC command.

Example Change the Resolution

CONF:RES 1560,MAX

Function: 2-wire ohms; range selected: 10kΩ;

 

MAX resolution: 1 Ω.

RES:RES 10E-03

Set resolution to 10mΩ.

READ?

Place multimeter in wait-for-trigger state and make

enter statement

measurements; send readings to output buffer.

Enter readings into computer.

RESistance:RESolution?

[SENSe:]RESistance:RESolution? [MINMAX] returns one of the following numbers to the output buffer.

The present resolution selected if MIN or MAX are not specified. Only the resolution values available on ranges set by the RANGe command are returned.

The resolution with the smallest value (i.e., the best resolution) for the selected range if MIN is specified.

The resolution with the largest value (i.e., the worst resolution) for the selected range if MAX is specified.