Triggering the Multimeter

This section discusses the multimeter’s trigger system and outlines the different triggering configurations and programming methods used to control the trigger system. Keep in mind that you do not have to program the trigger system to make measurements. You can avoid having to learn the information in this section by using the default trigger configuration set by MEASure and CONFigure commands. However, you will need the information in this section to take advantage of the flexibility of the

HP E1312A/E1412A trigger system when using the CONFigure command.

The multimeter’s trigger system synchronizes measurements with specified internal or external events. These events include software trigger commands, negative-going edges on the VXIbus trigger lines (TTLT0 - TTLT7), and negative-going pulses on the multimeter's external trigger (“Trig”) BNC connector. The trigger system also allows you to specify the number of triggers that will be accepted, the number of readings per trigger (sample count), and the delay between the trigger and each reading.

Figure 2-1 illustrates the multimeter's trigger system and the programming commands that control the trigger system. The multimeter operates in one of two trigger states. When you are configuring the multimeter for measurements, the multimeter must be in the idle state. After configuring the multimeter, the multimeter must be placed in the wait-for-triggerstate.

Figure 2-1. Multimeter Triggering Flow Chart

Chapter 2

HP E1312A/E1412A Multimeter Application Information 45

Page 45
Image 45
HP WaterSkis E1412A, WaterSkis E1312A manual Triggering the Multimeter