:LFRequency?

CALibration:LFRequency? queries the line frequency setting.

Comments This command returns +50 for line frequency set to 400 because 400 is an even multiple of 50.

Example Query the Line Frequency Setting

CAL:LFR?

Query the line frequency.

:SECure:CODE

CALibration:SECure:CODE <new code> enters a new calibration security code. To change the security code, first unsecure the multimeter using the old security code with CAL:SEC:STAT OFF, <old code>. Then, enter the new code. The calibration security code may contain up to 12 characters. The security code is stored in non-volatile memory.

Comments The security code is set to “ HP_E1412” for C-size (or “ HP_E1312” for B-size) when the multimeter is shipped from the factory. The security code is stored in non-volatile memory, and does not change when power has been off or after a remote interface reset.

The security code <new code> can contain up to 12 alphanumeric characters. The first character must be a letter. The remaining characters can be letters or numbers or an underscore. You do not have to use all 12 characters but the first character must be a letter.

If you forget or lose the active security code, you can disable the security feature by adding a jumper inside the multimeter (see Chapter 5 in the Service Manual). You then enter a new code and remove the jumper.

Example Enter a New Calibration Security Code

CAL:SEC:STAT OFF, HP_E1412

Unsecure with the old code.

CAL:SEC:CODE the_new_code

Enter a new calibration code

 

(a maximum of 12 characters).

80 Multimeter Command Reference