DC Characteristics (continued)

Measuring Characteristics
DC Voltage

Measurement Method:

A/D Linearity:

Input Resistance:

Input Bias Current:

Input Terminals:

Input Protection:

Resistance

Measurement Method:

Max. Lead Resistance:

(4-wire ohms)

Input Protection:

DC Current

Shunt Resistor:

Input Protection:

DC:DC Ratio

Measurement Method: Input HI-LO Reference HI-LO Input to Reference

Continuously integrating, multi-slope III A/D converter. 0.0002% of reading + 0.0001% of range

0.1V, 1V, 10V ranges

Selectable 10MΩ or >10GΩ

100V, 300V ranges

10MΩ ± 1%

<30pA at 25 °C

 

Copper alloy

 

300V on all ranges

 

Selectable 4-wire or 2-wire ohms. Current source referenced to LO input.

10% of range per lead for 100 Ω and 1kΩ ranges. 1kΩ per lead on all other ranges.

300V on all ranges

0.1 Ω for 1A and 3A. 5Ω for 10mA and 100mA

Externally accessible 3.15A, 250V, Class H fuse (see note at the bottom of the

AC Measuring Characteristics page describing class H fuses)

Input HI-LO/Reference HI-LO (Reference = Ω4W Sense terminals) 100mV to 300V ranges

100mV to 10V ranges (autoranged) Reference LO to Input LO voltage <2V Reference HI to Input LO voltage <12V

Measurement Noise Rejection

DC CMRR: 140 dB [5]

Normal Mode Rejection [6]

Integration Time
60 Hz (50 Hz)

 

 

100 PLC, 1.67s (2s)

60dB [7]

10 PLC, 167ms (200ms)

60dB [7]

1 PLC, 16.7ms (20ms)

60dB [7]

<1 PLC

0dB

[1]Specifications are for 1-hour warm-up at an integration time of 100 PLCs.

[2]Relative to calibration standards.

[3]20% overrange on all ranges, except 300Vdc and 3A range which have 1% overrange.

[4]Specifications are for 4-wire ohms function, or 2-wire ohms using Math Null. Without Math Null, add 0.2Ω additional error in 2-wire ohms function.

172 HP E1312A and HP E1412A Multimeter Specifications

Appendix A