:PERiod

CONFigure:PERiod [<range>MINMAXDEFAUTO

[,<resolution>MINMAXDEF]] selects the period function and allows you to specify range and resolution.

Parameters

 

 

 

 

Parameter Name

Parameter Type

Range of Values

Default Units

 

 

 

 

<range>

numeric

3.33E-01

Sec

 

 

 

 

<resolution>

numeric

3.33E-05 3.33E-06 3.33E-07

Sec

 

 

 

 

 

 

 

 

Comments The period function uses one “range” for all inputs between 0.33 seconds and

3.3 μSec. A period measurement will return “ 0” if no input is applied.

Range and resolution settings are listed below for the MIN, MAX, DEF and AUTO parameters and the settings after a module reset (*RST).

 

 

 

PARAMETER

RANGE

RESOLUTION

 

 

 

 

 

 

MIN

3.33E-01

3.33E-07

 

 

 

MAX

3.33E-01

3.33E-05

 

 

 

DEFAUTO and module

3.33E-01

3.33E-06

reset (*RST)

 

 

 

 

 

 

 

 

Chapter 3