AC Characteristics (continued)

Operating Characteristics

 

 

Function

 

Digits

Readings/sec

AC Filter

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

6 1/2

1

 

Slow (3Hz)

 

 

 

 

 

 

 

(per 7 seconds) [8]

(7 sec settling time)

 

 

 

 

 

 

 

 

 

 

 

 

 

ACV and ACI

 

6 1/2

1

[8]

Medium (20Hz)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

6 1/2

1.6

[8],[9]

Fast (200Hz)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

6 1/2

50

[10]

Fast (200Hz)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

System Speeds [10], [11]

 

 

 

 

 

 

 

Function or Range Change:5/sec

 

 

 

 

 

Autorange Time:

<0.8 sec

 

 

 

 

ASCII readings to HP-IB:50/sec

 

 

 

 

 

Max. Internal Trigger Rate:

50/sec

 

 

 

 

 

Max. External Trigger Rate to Memory:

50/sec

 

 

 

 

 

[1]Specifications are for 1-hour warm-up at 100 PLC integration time, 3 Hz (Slow) ac filter, sinewave input.

[2]Relative to calibration standards.

[3]20% overrange on all AC ranges, except 300V and 3A ranges which have 1% overrange.

[4]Values in the AC Characteristics Accuracy Specifications table are for sinewave inputs >5% of range (>15% of range for 300 VAC). For smaller inputs, add an additional error to the value in the table as follows:

 

 

 

 

 

 

 

 

 

 

 

 

 

Additional Error Specifications to % Range Table Value for Sinewave Inputs:

 

 

 

 

 

 

 

 

 

 

 

1% to 5% of range

3% to 15% of range

 

 

Function

Range

<50kHz

 

>50kHz

<50kHz

 

>50kHz

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

100mV to 100V

add 0.1 to

 

add 0.13 to

 

 

 

 

 

 

 

% of range

 

% of range

 

 

 

 

 

True RMS

 

 

 

 

 

 

 

 

 

300V

 

 

 

add 0.3 to

 

add 0.4 to

 

 

AC Voltage

 

 

 

 

 

 

 

 

 

 

% of range

 

% of range

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

True RMS

1A and 3A

add 0.1 to

 

 

 

 

 

 

 

AC Current

 

% of range

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

[5]300Vac range limited to 50kHz. For frequencies >50 kHz, signals must be <1.5 x 107 Volt-Hz.

[6]For frequencies below 100Hz, 3Hz (Slow) AC filter specified for sinewave input only.

[7]For 1kΩ unbalance in LO lead.

[8]Maximum reading rates for 0.01% of ac step additional error. Additional settling delay required when input dc level varies.

[9]For External Trigger or remote operation using default settling delay (Delay Auto).

[10]Maximum useful limit with default settling delays defeated.

[11]Speeds are for 0.02 PLC integration time, Delay 0, and 200Hz (Fast) ac filter.

[12]100Vac and 300Vac ranges may latch up the module or system mainframe if you drive the LO terminal with a high voltage, high frequency input. Only drive the HI terminal when measuring ac voltages.

176 HP E1312A and HP E1412A Multimeter Specifications

Appendix A