:SECure:STATe

CALibration:SECure:STATe OFFON, <code> unsecures or secures the multimeter for calibration. The calibration code must be the code set by the CAL:SEC:CODE command. The state is stored in non-volatile memory.

Parameters

 

 

 

 

 

 

 

 

Parameter Name

Parameter Type

Range of Values

Default Units

 

 

 

 

 

 

 

 

OFFON

boolean

OFF 0 ON 1

none

 

 

 

 

 

 

 

 

<code>

discrete

up to 12 characters

none

 

 

 

 

set by CAL:SEC:CODE

 

 

 

 

 

 

 

 

 

 

 

 

 

Comments

You can substitute decimal values for the OFF (“ 0”) and ON (“ 1”) parameters.

The multimeter calibration is secured when shipped from the factory. The security code is set to “ HP_E1412” (or “ HP_E1312” for B-size).

*RST does not change the state.

Example Set the Calibration State to Unsecured

CAL:SEC:STAT OFF, HP_E1412 Unsecure multimeter calibration.

:SECure:STATe?

CALibration:SECure:STATe? returns a “ 1” or “ 0” to show whether the calibration security state is enabled (1) or disabled (0). The number is sent to the output buffer.

Example Query the Calibration Security State

CAL:SEC:STAT?

Query multimeter calibration security state.

enter statement

Enter value into computer.

:STRing

CALibration:STRing <quoted string> allows you to record calibration information about your multimeter while CAL:SEC:STAT is OFF. For example, you can store information such as the last calibration date and/or the next calibration due date. The calibration message can contain up to 40 characters. Characters in excess of 40 are truncated and no error is generated. The string is stored in non-volatile memory.

Parameters

Parameter Name

Parameter Type

Range of Values

Default Units

 

 

 

 

<quoted string>

discrete

alphanumeric

none

 

 

 

 

 

 

 

 

Comments The calibration message can contain up to 40 characters.

Calibration security state must be OFF to store a string.

The calibration message is stored in non-volatile memory and does not change when power has been off or after a remote interface reset.