FRESistance:RANGe:AUTO

[SENSe:]FRESistance:RANGe:AUTO <mode> enables or disables the autorange

function for 4-wire resistance measurements.

Parameters

 

 

 

 

 

 

Parameter Name

Parameter Type

Range of Values

Default Units

 

 

 

 

 

 

<mode>

boolean

OFF0ON1

none

 

 

 

 

 

 

 

 

 

 

Comments

You can substitute decimal values for the OFF (“ 0”) and ON (“ 1”) parameters.

When autoranging is ON, the multimeter samples the input before each measurement and selects the appropriate range.

If you explicitly select a range using FRESistance:RANGe, autoranging is turned OFF.

Example Put 4-wire Resistance Measurements in the Autorange Mode

FRES:RANG:AUTO ON

Autorange is turned on for 4-wire ohms

 

measurements.

FRESistance:RANGe:AUTO?

[SENSe:]FRESistance:RANGe:AUTO? returns a number to show whether the autorange mode is enabled or disabled: “ 1” = ON, “ 0” = OFF. The number is sent to the output buffer.

Example Query the Autorange Mode

FRES:RANG:AUTO OFF

Disable autorange.

FRES:RANG:AUTO?

Query multimeter to return autorange mode.