Default Trigger Delays for DC Voltage and DC Current (all ranges):

 

Integration Time

 

 

Trigger Delay

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

NPLC ³1

 

 

1.5ms

 

 

 

 

 

 

 

 

 

NPLC <1

 

 

1.0ms

 

 

 

 

 

 

 

 

Default Trigger Delays for 2-Wire and 4-Wire Resistance:

 

 

 

 

 

 

 

 

 

Range

 

Trigger Delay

Trigger Delay

 

 

 

 

(For NPLC ³1)

(For NPLC <1)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

100W

 

1.5ms

1.0ms

 

 

 

 

 

 

 

 

1kW

 

1.5ms

1.0ms

 

 

 

 

 

 

 

 

10kW

 

1.5ms

1.0ms

 

 

 

 

 

 

 

 

100kW

 

1.5ms

1.0ms

 

 

 

 

 

 

 

 

1MW

 

1.5ms

10ms

 

 

 

 

 

 

 

 

10MW

 

100ms

100ms

 

 

 

 

 

 

 

 

100MW

 

100ms

100ms

 

 

 

 

 

 

 

 

Default Trigger Delays for AC Voltage and AC Current (all ranges):

 

AC Filter

Trigger Delay

 

 

 

 

 

 

 

 

 

 

3Hz - 300kHz filter (Slow)

7.0sec

 

 

 

 

 

 

20Hz - 300kHz filter (Medium)

1.0sec

 

 

 

 

 

 

200Hz - 300kHz filter (Fast)

600ms

 

 

 

 

 

Default Trigger Delay for Frequency and Period:

1.0s

:DELay:AUTO?

TRIGger:DELay:AUTO? returns a number to show whether the automatic trigger delay mode is on or off: “ 1” = ON, “ 0” = OFF. The number is sent to the output buffer.

Example Query the Trigger Delay Mode

TRIG:DEL:AUTO OFF

Disable automatic trigger delay.

TRIG:DEL:AUTO?

Query multimeter to return trigger delay mode.

enter statement

Enter value into computer.

Chapter 3

Multimeter Command Reference 159

Page 159
Image 159
HP WaterSkis E1412A, WaterSkis E1312A manual DELayAUTO?, Example Query the Trigger Delay Mode