621AC rms full scale failed

622Frequency counter failed

623Cannot calibrate precharge

625I/O processor does not respond

626I/O processor failed self-test

Calibration Errors The following errors indicate failures that may occur during a calibration. The most common errors have descriptions here. Refer to the HP E1312A and HP E1412A Service Manual for more information on the other errors.

701Cal security disabled by jumper

The calibration security feature has been disabled with a jumper inside the multimeter. When applicable, this error will occur at power-on to warn you that the multimeter is unsecured.

702Cal secured

The multimeter is secured against calibration.

703Invalid secure code

An invalid calibration security code was received when attempting to unsecure or secure the multimeter. You must use the same security code to unsecure the multimeter as was used to secure it, and vice versa. The security code may contain up to 12 alphanumeric characters. The first character must be a letter.

704Secure code too long

A security code was received which contained more than 12 characters.

705Cal aborted

A calibration in progress is aborted when you send a device clear to the multimeter.

706Cal value out of range

The specified calibration value (CAL:VALue) is invalid for the present function and range.

707Cal signal measurement out of range

The specified calibration value (CAL:VALue) does not match the signal applied to the multimeter.

708Cal signal frequency out of range

The input signal frequency for an ac calibration does not match the required input frequency for calibration.