:COUNt?

TRIGger:COUNt? [MINMAX] returns one of the following numbers to the output buffer:

The present trigger count (1 through 50,000) if MIN or MAX are not specified.

The minimum trigger count available (1) if MIN is specified.

The maximum trigger count available (50,000) if MAX is specified.

Example Query the Trigger Count

TRIG:COUN 10

Multimeter will accept 10 triggers.

TRIG:COUN?

Query multimeter to return trigger count.

enter statement

Enter value into computer.

:DELay

TRIGger:DELay <seconds> sets the delay time between receipt of the trigger and the start of the measurement. NOTE: This delay also occurs between each sample when SAMP:COUN > 1. See page 45 for a triggering process diagram.

Parameters

 

 

 

 

Parameter Name

Parameter Type

Range of Values

Default Units

 

 

 

 

<seconds>

numeric

0 through 3600MINMAX

seconds

 

 

 

 

 

 

 

 

Comments MIN selects the minimum delay of 0 seconds for all functions. MAX selects the maximum delay of 3600 seconds for all functions.

The trigger delay is inserted between the trigger and each measurement.

If a trigger delay is specified using the TRIG:DEL <period>,

TRIGger:DELay:AUTO is turned OFF.

The multimeter selects an automatic delay if you do not specify a trigger delay (see the TRIGger:DELay:AUTO command on page 158).

*RST Condition: TRIGger:DELay:AUTO ON

Example Set the Trigger Delay

TRIG:DEL .002

Wait 2ms between trigger and start of

 

measurement.