HP WaterSkis E1312A VOLTageACRESolution?, RST Condition 1E-04 Example Change the Resolution

Models: WaterSkis E1412A WaterSkis E1312A

1 212
Download 212 pages 21.44 Kb
Page 146
Image 146

VOLTage:AC:RESolution

[SENSe:]VOLTage:AC:RESolution <resolution> selects the resolution for AC voltage measurements. See Table 3-4 on page 71 to avoid a settings conflict error when specifying resolution.

Parameters

 

 

 

 

Parameter Name

Parameter Type

Range of Values

Default Units

 

 

 

 

<resolution>

numeric

resolutionMINMAX

volts

 

 

 

 

 

 

 

 

Comments MINimum selects the best resolution (the smallest value) for the selected range. MAXimum selects the worst resolution (the largest value) for the selected range.

You must select a range using VOLTage:AC:RANGe before specifying resolution. Also, only specify a resolution when making measurements on a fixed range. Otherwise, the resolution will change to correspond with the range selected during autoranging.

The VOLT:AC:RES command overrides the resolution setting from a previous

CONFigure:VOLT:AC:RES command.

Related Commands: CONFigure, VOLTage:DC:RESolution

*RST Condition: 1E-04

Example Change the Resolution

CONF:VOLT:AC 6.25,MAX

Function: DC volts; range selected: 10V;

 

MAX resolution.

VOLT:AC:RANG 0.95

Range selected: 1.0V; MAX resolution: 100 μV.

VOLT:AC:RES 10E-06

Set resolution to 10 μV.

READ?

Place multimeter in wait-for-trigger state and make

 

measurements; send readings to output buffer.

VOLTage:AC:RESolution?

[SENSe:]VOLTage:AC:RESolution? [MIN MAX] returns one of the following numbers to the output buffer.

The present resolution selected if MIN or MAX is not specified. Only the resolution values available on ranges set by the RESolution command are returned.

The resolution with the smallest value (i.e., the best resolution) for the selected range if MIN is specified.

The resolution with the largest value (i.e., the worst resolution) for the selected range if MAX is specified.

Example Query the Resolution

VOLT:AC:RANG 100E-03

Set range to 0.1 volts.

VOLT:AC:RES 1.0E-07

Set resolution to 0.1 μV.

VOLT:AC:RES?

Query multimeter to return the present

enter statement

resolution.

Enter value into computer.

146 Multimeter Command Reference