:SOURce

TRIGger:SOURce <source> configures the trigger system to respond to the specified source. The following sources are available:

BUS: Group Execute Trigger (GET) bus command or *TRG common command.

EXT: The multimeter’s External Trigger BNC connector (triggers on the negative or falling edge of the ±5V TTL input signal).

IMMediate: The trigger system is always true.

TTLTrg0 - TTLTrg7: Trigger source is VXIbus trigger line 0 through 7.

NOTE: B-size controllers do not support VXIbus TTL triggers (e.g.,

HP E1306A Command Module, HP E1300/E1301A B-Size Mainframes).

Parameters

Parameter Name

Parameter Type

Range of Values

Default Units

 

 

 

 

<source>

discrete

BUSEXTIMMediate

none

 

 

TTLTrg0 through TTLTrg7

 

 

 

 

 

 

 

 

 

Comments The TRIGger:SOURce command only selects the trigger source. You must use the INITiate command to place the multimeter in the wait-for-trigger state. (The MEASure command automatically executes an INITiate command.)

TRIGger:SOURce EXT uses the multimeter's front panel “Trig” BNC connector as the trigger source. The multimeter triggers on the falling (negative-going) edge of a ±5V TTL input signal; (maximum input is +5V to the front panel BNC connector).

TRIGger:IMMediate causes a trigger to occur immediately provided the multimeter is placed in the wait-for-trigger state using INITiate, READ? or

MEAS?.

When a Group Execute Trigger (GET) bus command or *TRG common command is executed and the multimeter is not in the wait-for-trigger state, the “Trigger ignored” error is generated.

The CONFigure and MEASure command subsystems automatically set the trigger source to TRIG:SOUR IMM.

The READ? command cannot be used if the trigger source is

TRIG:SOUR BUS.

Related Commands: INITiate, READ?, MEAS?

*RST Condition: TRIG:SOUR IMM

Example Set the Sample Source

CONF:VOLT:DC

Function: DC voltage.

TRIG:SOUR EXT

Trigger source is external BNC on multimeter

TRIG:COUN 10

front panel.

Multimeter will accept 10 external triggers.

READ?

Place multimeter in wait-for-trigger state;

 

make measurements when external trigger is

enter statement

received; send readings to output buffer.

Enter readings into computer.

160 Multimeter Command Reference

Chapter 3

Page 160
Image 160
HP WaterSkis E1312A, WaterSkis E1412A manual SOURce, Related Commands INITiate, READ?, MEAS?, Example Set the Sample Source