[:VOLTage[:DC]]:RATio

CONFigure[:VOLTage[:DC]]:RATio [<range>MINMAXDEFAUTO [,<resolution>MINMAXDEF]] configures the multimeter for dc:dc ratio measurements with the specified range and resolution.

dc signal voltage

Hi and LO input

DC:DC RATIO = dc------------------------------------------------reference voltage =

------------------------------------------------------

Sense HI and LO input

The ratio is calculated from the voltage applied to the HI and LO input terminals divided by the reference voltage applied to the “Sense” HI and LO terminals. Autoranging is automatically selected for the reference voltage measurement on the “Sense” HI and LO terminals. The specified range in the command applies to the signal connected to the HI and LO input terminals.

Note Autorange on the “Sense” terminals is from 100mV to 10V range only. Maximum voltage you can apply to the “Sense” terminals is 10V.

Parameters

Parameter Name

Parameter Type

Range of Values

Default Units

 

 

 

 

<range>

numeric

100mV1V10V100V300V

volts

(HI-LO input)

 

MINMAXDEFAUTO

 

 

 

 

 

<resolution>

numeric

resolution MINMAXDEF

volts

 

 

 

 

 

 

 

 

Comments To select a standard measurement range, specify range as the input signal’s maximum expected voltage. The multimeter then selects the correct range to accept the input.

The AUTO or DEFault option for the range parameter enables autorange. The DEFault option for resolution defaults the integration time to 10 PLC.

The MIN and MAX parameters select the minimum or maximum values for range and resolution:

For range: MIN = 100mV; MAX = 300V.

For resolution: MIN selects the best resolution (the smallest value) for the selected range. MAX selects the worst resolution (the largest value) for the selected range.