AC Characteristics (continued)

Measuring Characteristics

Measurement Noise Rejection [7]

AC CMRR

70 dB

True RMS AC Voltage

Measurement Method: AC-coupled True RMS - measures the ac component of the input with up to 300Vdc of bias on any range (max AC + DC = 300Vrms).

Crest Factor:

Maximum 5:1 at full scale

AC Filter Bandwidths:

3Hz-300kHz (Slow filter)

 

20Hz-300kHz (Medium filter)

 

200Hz-300kHz (Fast filter)

Input Impedance:

1MΩ ± 2%, in parallel with 100pF

Input Protection:

300Vrms all ranges

True RMS AC Current

Measurement Method: Direct coupled to the fuse and shunt. AC-coupled True RMS measurement (measures the ac component only).

Shunt Resistor:

0.1 Ω for 1A and 3A ranges

Burden Voltage:

1A range: <1Vrms

 

3A range: < 2Vrms

Input Protection:

Externally accessible 3.15A, 250V, Class H fuse

 

Class H fuses are fuses with a high interrupt rating which defines a fuse’s ability

 

to safely interrupt and clear short circuits. Replace the fuse with HP part number

 

2110-0957, (3.15A, 250V, 5.0 mm diameter, 20.0 mm long) or use Cooper Industries

 

Inc. fuse part number GDA-3.15.