Replacing the pre- separation pad

1.Remove the separator module as previously directed.

2.Locate the pre-separation pad on the separator module.

3.Hold the separator module in both hands and find the two pre- separation pad side tabs which protrude slightly from the back of the separator module.

4.Push on the pre-separation pad side tabs on the back until the pre- separation pad protrudes slightly from the front of the separator module.

5.Grasp the pre-separation pad and pull it out of the separator module.

6.Discard the worn pre-separation pad.

7.Align the side tabs of a new pre-separation pad with the slots on the separator module.

8.Push the pre-separation pad firmly in the slots until the pre- separation pad flap rests lightly against the separator module tires.

9.Reinstall the separator module.

10.Close the scanner cover.