4.0 PROGRAMMING

When prior to operating the OM-550 it must first be programmed. This section covers all of the programming options available. All programmed settings but the time & date are stored in EEPROM and retained even if the batteries are removed.

Programming outline: 1. Select Channel (1-6)

If Channel (1-5)

Set Channel Type (J,K,T,E,V,Skip)

If Channel Type (J,K,T, Or E)

(Return To Select Channel)

If Channel V (Voltage) Set Mx And B

Set Engineering Units

Set Decimal Position

(Return To Select Channel)

If Channel Type Skip

(Return To Select Channel)

If Channel 6 (Math)

If Channel Type Skip

(Return To Select Channel)

If Not Channel Type Skip

Set Math Function ( +, - , /, X )

Select First Operand Channel

Select Second Operand Channel

Set Engineering Units

Set Decimal Fix Position

(Return To Select Channel)

2.Set Log Interval

3.Select Activity at Full Cache

4.Set Date

5.Set Time

6.Select Temperature Units (F/C)

7.Select 60/50 Hz A/D Reject

8.Select Calibration Mode

If Normal Mode Selected Set Tadjust Programming Done

If Calibration Mode Selected Programming Done

When operating the OM-550 and you are unsure of the programmed settings, a Manual Reset can be performed to force it to a standard configuration (refer to section APPENDIX A: CALIBRATION for more information).