Calibrating I/O

Calibrating Analog Channels

Analog Input can be calibrated for Analog and Temperature IOLAN models.

Select the channel you want to calibrate. This example uses an A4 model that has channel A1 set to Current with a Range of 0 to 20mA.

If you have not disabled confirmation messages (Tools, Options in DeviceManager only), you will get prompted to verify channel calibration.

Click Yes to proceed with calibration. You are now prompted to apply 0 mA to the positive (+) and negative (-) terminals. Once that is done, click Yes to proceed.

You are now prompted to apply 20 mA to the positive (+) and negative (-) terminals. Once that is done, click Yes to proceed.

Once calibration is successfully completed, click OK to finish the process.

Resetting Calibration Data

You can reset the I/O channels calibrations to the factory calibrations in DeviceManager by selecting Tools, I/O Channels, Reset Calibrate Data or in WebManager by selecting Administration, Reset, I/O Calibration.