Alarm message | Possible cause | |
Pump X Option Output Conflict | An assigned station has closed loop enabled and the pump has either | |
implosion with sealing valve or Local Alarm enabled. | ||
| ||
Purge Valve X Failure | Patch cord is cut or disconnected | |
Solenoid has failed. | ||
| ||
| ArmorBlock is not installed properly. | |
Purge Valve X Not Present | ArmorBlock is not addressed correctly. | |
ArmorBlock is defective. | ||
| ||
| ArmorBlock base is defective. | |
Purge Valve X Power Failure | Patch cord is cut or disconnected. | |
Solenoid has failed. | ||
| ||
Receiver Level | Pick up wand is not positioned properly. | |
| Leak in vacuum line. | |
RPV Failure | Patch cord is cut or disconnected. | |
Solenoid has failed. | ||
| ||
Sequence Valve Failure | Patch cord is cut or disconnected. | |
Solenoid has failed. | ||
| ||
Slide Gate Failure | Patch cord is cut or disconnected. | |
Solenoid has failed. | ||
|
| Chapter 7: Troubleshooting | 73 |