8.Working with the Instrument

A few rules:

Always begin at a low extraction level, then slowly increase it.

Do not use high extraction levels unless absolutely necessary.

Different specimen sizes require different extraction levels.

The faster the trimming or sectioning speed, the lower the extraction

level

should be.

The larger and/or thicker the specimen to be trimmed, the lower the extraction value.

When section specimens with a diameter of 0.5 cm, the anti-roll guide stretches the section adequately. For larger specimens, we recommend using the vacuum function.

Trimming, 1 x 1cm specimen

 

Trimming thickness

Speed

Extract. value

 

 

 

10μm

40%

1.5

10μm

60%

1.5

10μm

80%

1

10μm

100%

1

 

 

 

 

 

 

20μm

40%

2

20μm

60%

2

20μm

80%

1.5

20μm

100%

1

 

 

 

 

 

 

30μm

40%

2.5

30μm

60%

2.5

30μm

80%

2

30μm

100%

1.5

 

 

 

 

 

 

40μm

40%

3

40μm

60%

3

40μm

80%

2.5

40μm

100%

2

 

 

 

Trimming, 1 x 1cm specimen

 

Trimming thickness

Speed

Extract. value

 

 

 

1μm - 5 μm

40%

2.5

 

60%

2

 

80%

1.5

 

100%

1.5

 

 

 

 

 

 

10μm

40%

2.5

 

60%

2

 

80%

2

 

100%

1.5

 

 

 

The values listed here are possible settings that depend on the tissue type, specimen size, section thick- ness and section speed, and are in- tended for orientation purposes only.

These settings can also vary when using different anti-roll plate!

46

Operating Manual V1.1 – 05/2007

Image 46
Contents Leica CM1950 Page Page Table of contents 12.2 11.112.1 12.3Qualification of Personnel Symbols Used in This Manual and Their MeaningImportant Information Designated UseGeneral Safety Instructions Safety and DesignUnpacking and Installation Opening the packaging Transport to the Installation Location Ramp assemblyRemoving the packaging Safety Features Microtome knivesKnife guard Securing/Locking the HandwheelCleaning, Disinfection Turning the Instrument Back On Handling Specimens Defrosting MaintenanceRemoval of the Microtome Technical Data Technical Data Refrigeration system CM1950, 50 Hz CM1950, 60 HzStandard Delivery Set of filters 5 piecesHandwheel, motorized General Overview Control Panel Fields and Cryostat Chamber Right side InstallationSite Requirements Left sideTransport to the Site Assembling the Handwheel Centering the Handwheel Optional Locking/Unlocking the HandwheelEmergency-stop function active, or Existing power system Connection to Power Supply SystemOnly for instruments sold in Japan Selecting the frequency Installation Inserting the Section Waste Tray Installing the Storage Systems Optional For reasonsShelf, Movable Optional Section waste tray with the a Cutout a facing the userInstalling the Heat Conductor Block, Stationary Optional Installation Mark for suction nozzle Instrument Controls Control Panel Fields on the CM1950 Control Panel Field12 o’clock Position 6 o’clock 20.0μm Retraction0μm 60.0μmMotion continues as long as the button is held Coarse feed functionsBrief press of the button retracts the specimen by 20μm Control panel 3 Motorized sectioning optional Sectioning modesDisinfection Duration 30min Duration 180 min Leica Cryofect is not available in all countriesWorking with the Instrument Turning the Instrument OnConfiguring the Parameters Working with the Instrument Working with the Instrument Manual defrosting of the specimen cooling Entering the trimming thickness Working with the Precooled Cryostat 8.4.1 Preparatory Work Trimming with Extraction Anti-roll Guide Installed Handwheel position 12 3 o’clock, valve open all the way Cutting with Extraction Anti-roll Guide installed Shut off VAC LED in VAC button goes out10μm 40% 60% 80% 100% 20μm 30μm 40μm SpeedSpeed Extract. value 1μm 5 μm 40% 60% 80% 100% 10μm Working with the Instrument Problems During Work TroubleshootingProblem Causes Remedies Cryochamber Sections tear Alternately Longer sufficient Temperature Selection Chart in minus C Tissue type Chamber temperatureOptional Accessories Ordering InformationCryofect, VPE Leica Cryofect 4x 350ml Lever 6 for the lateral shift When using low-profile blades, the rulerWith anti-roll guide Lever 7 for clamping the bladeBlade holder CE with finger rest for NB + BB Inserting the blade into knife holder CE Will result Cleaning the knife holder CE Daily cleaning Unscrew the bolts 35 of the pressure plateNot Visible Optional Accessories Anti-roll guide system Optional Accessories Specimen discs in various dimensions Optional Accessories Maintenance and Cleaning General MaintenanceChanging Fuses Replacement of the Lamp/UVC Lamp Installation of the new lamp NB Microtome knives must be in boxes This equipment has been cleaned and decontaminatedDecontamination Certificate master Yes No indicated belowBU Return Authorization Number Anne De Greef-Safft President Biosystems Division EC Declaration of ConformityWarranty and Service Peoples Republic of China