9. Troubleshooting

Problem

Causes

Remedies

 

Blunt knife.

Use different part of the knife.

 

Inappropriate section thickness.

Select correct section thickness.

 

Incorrect clearance angle.

Set correct angle.

 

Microtome not dried thoroughly

Dry microtome.

 

enough.

 

 

Dried specimen.

Prepare new specimen.

 

 

 

Tissue sticks to the anti-roll plate

Anti-roll plate is too warm or incorrectly positioned.

Fat on the corner or edge of anti-roll plate.

Anti-roll plate not correctly fixed. Rust on the knife.

Cool down anti-roll plate, or reposition correctly.

Remove fat from anti-roll plate.

Fix correctly.

Remove rust.

Flattened sections curl up when anti-roll plate is folded up

Anti-roll plate too warm.

Cool down anti-roll plate.

Sections tear

Temperature too low for the tissue cut.

Increase temperature and wait.

 

Blunt part, dirt, dust, frost or rust on

Remove cause.

 

the knife.

 

 

Top edge of the anti-roll plate

Replace anti-roll plate.

 

damaged.

 

 

Hard particles in the tissue.

- - -

 

Dirt on knife back.

Clean.

 

 

 

Cryostat non-operational

Power plug not properly connected.

 

Defective fuses, or circuit breaker has

 

triggered.

Check if properly connected. Replace fuses, or switch circuit breaker back on. If not possible, call technical service.

Specimen disc cannot be

Moisture on the underside cause the

removed

specimen to freeze to the freezing

 

shelf or specimen head.

Apply concentrated alcohol to the contact point.

50

Operating Manual V1.1 – 05/2007

Image 50
Contents Leica CM1950 Page Page Table of contents 12.2 11.112.1 12.3Qualification of Personnel Symbols Used in This Manual and Their MeaningImportant Information Designated UseUnpacking and Installation Safety and DesignGeneral Safety Instructions Opening the packaging Removing the packaging Ramp assemblyTransport to the Installation Location Safety Features Microtome knivesKnife guard Securing/Locking the HandwheelCleaning, Disinfection Turning the Instrument Back On Removal of the Microtome MaintenanceHandling Specimens Defrosting Technical Data Technical Data Refrigeration system CM1950, 50 Hz CM1950, 60 HzStandard Delivery Set of filters 5 piecesHandwheel, motorized General Overview Control Panel Fields and Cryostat Chamber Right side InstallationSite Requirements Left sideTransport to the Site Assembling the Handwheel Emergency-stop function active, or Locking/Unlocking the HandwheelCentering the Handwheel Optional Only for instruments sold in Japan Selecting the frequency Connection to Power Supply SystemExisting power system Installation Inserting the Section Waste Tray Installing the Storage Systems Optional For reasonsShelf, Movable Optional Section waste tray with the a Cutout a facing the userInstalling the Heat Conductor Block, Stationary Optional Installation Mark for suction nozzle Instrument Controls Control Panel Fields on the CM1950 Control Panel Field12 o’clock Position 6 o’clock 20.0μm Retraction0μm 60.0μmBrief press of the button retracts the specimen by 20μm Coarse feed functionsMotion continues as long as the button is held Control panel 3 Motorized sectioning optional Sectioning modesDisinfection Duration 30min Duration 180 min Leica Cryofect is not available in all countriesWorking with the Instrument Turning the Instrument OnConfiguring the Parameters Working with the Instrument Working with the Instrument Manual defrosting of the specimen cooling Entering the trimming thickness Working with the Precooled Cryostat 8.4.1 Preparatory Work Trimming with Extraction Anti-roll Guide Installed Handwheel position 12 3 o’clock, valve open all the way Cutting with Extraction Anti-roll Guide installed Shut off VAC LED in VAC button goes outSpeed Extract. value 1μm 5 μm 40% 60% 80% 100% 10μm Speed10μm 40% 60% 80% 100% 20μm 30μm 40μm Working with the Instrument Problem Causes Remedies TroubleshootingProblems During Work Cryochamber Sections tear Alternately Longer sufficient Temperature Selection Chart in minus C Tissue type Chamber temperatureOptional Accessories Ordering InformationCryofect, VPE Leica Cryofect 4x 350ml Lever 6 for the lateral shift When using low-profile blades, the rulerWith anti-roll guide Lever 7 for clamping the bladeBlade holder CE with finger rest for NB + BB Inserting the blade into knife holder CE Will result Cleaning the knife holder CE Daily cleaning Unscrew the bolts 35 of the pressure plateNot Visible Optional Accessories Anti-roll guide system Optional Accessories Specimen discs in various dimensions Optional Accessories Maintenance and Cleaning General MaintenanceChanging Fuses Replacement of the Lamp/UVC Lamp Installation of the new lamp NB Microtome knives must be in boxes This equipment has been cleaned and decontaminatedDecontamination Certificate master Yes No indicated belowBU Return Authorization Number Anne De Greef-Safft President Biosystems Division EC Declaration of ConformityWarranty and Service Peoples Republic of China