Leica CM1950 manual Trimming with Extraction Anti-roll Guide Installed

Page 43

8. Working with the Instrument

8.4.2 Trimming with Extraction – Anti-roll Guide Installed

E

Fig. 38

Remove the silicone stopper from the filter cover (and keep it in a safe place).

Insert the extraction hose with the black adapter.

Fold the anti-roll guide to the side and fasten the suction nozzle to the pressure plate – see mark Fig. 39a - (using 4 magnets on the rear side of the nozzle).

Fold the anti-roll guide back into position.

Fig. 39a

K B D

Fig. 39b

Fig. 39c

Ensure that the hose with the nozzle is not installed against its "natural" cur- vature on the pressure plate of the blade holder.

C

I

A

Fig. 40

The tension acting on the hose can be minimized by turning the red ring (Fig. 39, top right) clock- wise so that the suction nozzle presses against the pressure plate (I, Fig. 39).

Check that the anti-roll guide is parallel and correctly adjusted. Readjust if necessary (see page 58).

Activate trimming mode.

Select the trimming thickness.

Switch on the VAC and select a low extraction level (between 1 and 2).

Start trimming by manually moving the handwheel or press the RUN/STOP and ENABLE keys simultaneously to start motorized sectioning.

Leica CM1950 – Cryostat

43

Image 43
Contents Leica CM1950 Page Page Table of contents 12.3 11.112.1 12.2Designated Use Symbols Used in This Manual and Their MeaningImportant Information Qualification of PersonnelGeneral Safety Instructions Safety and DesignUnpacking and Installation Opening the packaging Transport to the Installation Location Ramp assemblyRemoving the packaging Microtome knives Safety FeaturesSecuring/Locking the Handwheel Knife guardCleaning, Disinfection Turning the Instrument Back On Handling Specimens Defrosting MaintenanceRemoval of the Microtome Technical Data Refrigeration system CM1950, 50 Hz CM1950, 60 Hz Technical DataSet of filters 5 pieces Standard DeliveryHandwheel, motorized General Overview Control Panel Fields and Cryostat Chamber Left side InstallationSite Requirements Right sideTransport to the Site Assembling the Handwheel Centering the Handwheel Optional Locking/Unlocking the HandwheelEmergency-stop function active, or Existing power system Connection to Power Supply SystemOnly for instruments sold in Japan Selecting the frequency Installation Section waste tray with the a Cutout a facing the user Installing the Storage Systems Optional For reasonsShelf, Movable Optional Inserting the Section Waste TrayInstalling the Heat Conductor Block, Stationary Optional Installation Mark for suction nozzle Control Panel Fields on the CM1950 Control Panel Field Instrument Controls12 o’clock Position 6 o’clock 60.0μm Retraction0μm 20.0μmMotion continues as long as the button is held Coarse feed functionsBrief press of the button retracts the specimen by 20μm Sectioning modes Control panel 3 Motorized sectioning optionalLeica Cryofect is not available in all countries Disinfection Duration 30min Duration 180 minTurning the Instrument On Working with the InstrumentConfiguring the Parameters Working with the Instrument Working with the Instrument Manual defrosting of the specimen cooling Entering the trimming thickness Working with the Precooled Cryostat 8.4.1 Preparatory Work Trimming with Extraction Anti-roll Guide Installed Handwheel position 12 3 o’clock, valve open all the way Shut off VAC LED in VAC button goes out Cutting with Extraction Anti-roll Guide installed10μm 40% 60% 80% 100% 20μm 30μm 40μm SpeedSpeed Extract. value 1μm 5 μm 40% 60% 80% 100% 10μm Working with the Instrument Problems During Work TroubleshootingProblem Causes Remedies Cryochamber Sections tear Alternately Longer sufficient Tissue type Chamber temperature Temperature Selection Chart in minus COrdering Information Optional AccessoriesCryofect, VPE Leica Cryofect 4x 350ml Lever 7 for clamping the blade When using low-profile blades, the rulerWith anti-roll guide Lever 6 for the lateral shiftBlade holder CE with finger rest for NB + BB Inserting the blade into knife holder CE Will result Unscrew the bolts 35 of the pressure plate Cleaning the knife holder CE Daily cleaningNot Visible Optional Accessories Anti-roll guide system Optional Accessories Specimen discs in various dimensions Optional Accessories General Maintenance Maintenance and CleaningChanging Fuses Replacement of the Lamp/UVC Lamp Installation of the new lamp Yes No indicated below This equipment has been cleaned and decontaminatedDecontamination Certificate master NB Microtome knives must be in boxesBU Return Authorization Number EC Declaration of Conformity Anne De Greef-Safft President Biosystems DivisionWarranty and Service Peoples Republic of China