Leica CM1950 manual Troubleshooting, Problems During Work, Problem Causes Remedies

Page 48

9.Troubleshooting

9.1Problems During Work

Problem

Causes

Remedies

 

 

 

Frost on chamber walls and microtome

Cryostat is exposed to air currents (open windows and doors, air conditioning).

Frost built-up by breathing into the cryochamber.

Change place of installation for the cryostat.

Wear mouth protection.

Sections smear

Specimen not cold enough.

 

Anti-roll plate not yet cold enough,

 

thus warming the section.

Select lower temperature. Wait until knife and/or anti-roll plate have reached chamber temperature.

Sections splinter

Specimen too cold.

Select higher temperature.

 

 

 

Sections not properly

Static electricity/air currents.

Remove cause.

flattened

Specimen not cold enough.

Select lower temperature.

 

Large area specimen.

Trim the specimen parallel,

 

 

increase section thickness.

 

Anti-roll plate poorly positioned.

Reposition anti-roll plate.

 

Anti-roll plate poorly aligned with knife

Align correctly.

 

edge.

 

 

Incorrect clearance angle.

Set correct angle.

 

Blunt knife.

Use different part of the knife.

 

 

 

Sections not properly flattened despite correct temperature and correctly aligned anti-roll plate

Dirt on knife and/or anti-roll plate. Edge of anti-roll plate damaged. Blunt knife.

Clean with dry cloth or brush. Replace anti-roll plate.

Use different part of the knife.

Sections curl on anti-roll

Anti-roll plate does not protrude far

Readjust correctly.

plate

enough beyond the knife edge.

 

 

 

 

Scraping noise during

Anti-roll plate protrudes too far beyond

Readjust correctly.

sectioning and specimen

the knife edge and is scraping against

 

return movement

the specimen.

 

48

Operating Manual V1.1 – 05/2007

Image 48
Contents Leica CM1950 Page Page Table of contents 11.1 12.112.2 12.3Symbols Used in This Manual and Their Meaning Important InformationQualification of Personnel Designated UseSafety and Design General Safety InstructionsUnpacking and Installation Opening the packaging Ramp assembly Transport to the Installation LocationRemoving the packaging Safety Features Microtome knivesKnife guard Securing/Locking the HandwheelCleaning, Disinfection Turning the Instrument Back On Maintenance Handling Specimens DefrostingRemoval of the Microtome Technical Data Technical Data Refrigeration system CM1950, 50 Hz CM1950, 60 HzStandard Delivery Set of filters 5 piecesHandwheel, motorized General Overview Control Panel Fields and Cryostat Chamber Installation Site RequirementsRight side Left sideTransport to the Site Assembling the Handwheel Locking/Unlocking the Handwheel Centering the Handwheel OptionalEmergency-stop function active, or Connection to Power Supply System Existing power systemOnly for instruments sold in Japan Selecting the frequency Installation Installing the Storage Systems Optional For reasons Shelf, Movable OptionalInserting the Section Waste Tray Section waste tray with the a Cutout a facing the userInstalling the Heat Conductor Block, Stationary Optional Installation Mark for suction nozzle Instrument Controls Control Panel Fields on the CM1950 Control Panel Field12 o’clock Position 6 o’clock Retraction 0μm20.0μm 60.0μmCoarse feed functions Motion continues as long as the button is heldBrief press of the button retracts the specimen by 20μm Control panel 3 Motorized sectioning optional Sectioning modesDisinfection Duration 30min Duration 180 min Leica Cryofect is not available in all countriesWorking with the Instrument Turning the Instrument OnConfiguring the Parameters Working with the Instrument Working with the Instrument Manual defrosting of the specimen cooling Entering the trimming thickness Working with the Precooled Cryostat 8.4.1 Preparatory Work Trimming with Extraction Anti-roll Guide Installed Handwheel position 12 3 o’clock, valve open all the way Cutting with Extraction Anti-roll Guide installed Shut off VAC LED in VAC button goes outSpeed 10μm 40% 60% 80% 100% 20μm 30μm 40μmSpeed Extract. value 1μm 5 μm 40% 60% 80% 100% 10μm Working with the Instrument Troubleshooting Problems During WorkProblem Causes Remedies Cryochamber Sections tear Alternately Longer sufficient Temperature Selection Chart in minus C Tissue type Chamber temperatureOptional Accessories Ordering InformationCryofect, VPE Leica Cryofect 4x 350ml When using low-profile blades, the ruler With anti-roll guideLever 6 for the lateral shift Lever 7 for clamping the bladeBlade holder CE with finger rest for NB + BB Inserting the blade into knife holder CE Will result Cleaning the knife holder CE Daily cleaning Unscrew the bolts 35 of the pressure plateNot Visible Optional Accessories Anti-roll guide system Optional Accessories Specimen discs in various dimensions Optional Accessories Maintenance and Cleaning General MaintenanceChanging Fuses Replacement of the Lamp/UVC Lamp Installation of the new lamp This equipment has been cleaned and decontaminated Decontamination Certificate masterNB Microtome knives must be in boxes Yes No indicated belowBU Return Authorization Number Anne De Greef-Safft President Biosystems Division EC Declaration of ConformityWarranty and Service Peoples Republic of China